POTENTIAL FOR ELECTROMAGNETIC
INTERFERENCE
To avoid inaccurate results caused by electro-
magnetic interference between electrical and
electronic equipment, do not use the device
near a cell phone or microwave oven.
This device complies with Part 15 of the FCC
Rules. Operation is subject to the following two
conditions: (1) this device may not cause harm-
ful interference, and (2) this device must accept
any interference received, including interference
that may cause undesired operation.
NOTE: The manufacturer is not responsible for any radio or TV interfer-
ence caused by unauthorized modifications to this equipment. Such
modifications could void the user authority to operate the equipment.
NOTE: This equipment has been tested and found to comply with the
limits for a Class B digital device, pursuant to Part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harm-
ful interference in a residential installation. This equipment generates,
uses and can radiate radio frequency energy and, if not installed and
used in accordance with the instructions, may cause harmful interference
to radio communications. However, there is no guarantee that interference
will not occur in a particular installation. If this equipment does cause
harmful interference to radio or television reception, which can be deter-
mined by turning the equipment off and on, the user is encouraged to try
to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from
that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
27

TROUBLESHOOTING

If any abnormality arises during use, please check the following points
POSSIBLE
SYMPTOMS
CAUSES
Unit does not turn
Batteries have
on when the START/
run down.
STOP button is
pushed.
Battery polarities
have been positioned
incorrectly.
EE measurement error
Cuff has been placed
symbol shown on dis-
incorrectly.
play or the blood pres-
sure value is displayed
excessively low
Did you talk or move
(or high).
during measurement?
Shaking of the arm with
the cuff on.
E1 error symbol shown
Air circuit abnormality.
on display.
Cuff tube may not be
plugged into monitor
correctly.
E2 error symbol shown
Inflation pressure
on display.
exceeding 300 mmHg.
E3 error symbol shown
Error determining
on display.
measurement data.
EP error symbol shown
System Error.
on display.
Note: If the unit still does not work, contact HoMedics Consumer Relations.
Under no circumstance should you disassemble or attempt to repair the
unit by yourself. Contact information for HoMedics Consumer Relations
Department can be found on the warranty page.
CORRECTION
Replace them with
four new AA
alkaline batteries.
Re-insert the batteries
in the correct positions.
Wrap the cuff properly
so that it is positioned
correctly.
Measure again. Keep
arm steady during
measurement.
Check cuff connection.
Measure again.
Switch the unit off, then
measure again.
Measure again.
Measure again.