Teledyne 4700 Pocket Manual page 68

Refrigerated sampler
Hide thumbs Also See for 4700:
Table of Contents

Advertisement

4700 Refrigerated Sampler
• your programmed number of flow pulses
has been reached, or
• a disabled sampler becomes enabled
(section 3.3.6).
At each event, the sampler:
1. Resets the programmed flow or time pacing
interval.
2. Moves the distributor arm over the next
bottle. (This step does not occur for
samplers configured with the 20 liter
composite bottle.)
3. Rinses the suction line (see section 3.3.4).
4. Sends a three-second event mark signal
that begins at forward pumping. The
sampler deposits the programmed sample
volume into the bottle.
If programmed to deposit a sample volume
in more than one bottle, the sampler:
a. Moves the distributor arm over the next
bottle.
b. Deposits the programmed sample
volume into the bottle.
c. Repeats steps 4a and 4b until the
programmed number of bottles per event
is reached.
5. Rinses the suction line (see section 3.3.4).
3-20