Endress+Hauser Nanomass Gas Density Operating Instructions Manual

Mems coriolis density meter
Table of Contents

Advertisement

Quick Links

BA01449D/06/EN/03.16
71311429
Valid as of version
V 3.00.9 (device software)
V 3.01.3 (Nanomass Communication)
Products

Operating Instructions

Nanomass Gas Density
MEMS Coriolis density meter
Solutions
-
E S
C
P o
w e
r
+
E
E r r
o r
Services

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Nanomass Gas Density

  • Page 1: Operating Instructions

    Products Solutions Services BA01449D/06/EN/03.16 71311429 Valid as of version V 3.00.9 (device software) V 3.01.3 (Nanomass Communication) Operating Instructions Nanomass Gas Density MEMS Coriolis density meter E r r...
  • Page 2 Operating Instructions that are specific to working procedures. • The manufacturer reserves the right to modify technical data without prior notice. Your Endress+Hauser Sales Center will supply you with current information and updates to these Operating Instructions. Endress+Hauser...
  • Page 3: Table Of Contents

    13.2 Measuring and test equipment ....59 13.3 Endress+Hauser services ....60 Electrical connection .
  • Page 4 Nanomass Gas Density 16.3 Input ........65 16.4 Output .
  • Page 5: Notes On The Operating Instructions

    Nanomass Gas Density Notes on the Operating Instructions Notes on the Operating Instructions Document function These Operating Instructions contain all the information that is required in the various phases of the life cycle of the device: from product identification, incoming acceptance and storage, to mounting, connection, operation and commissioning through to troubleshooting, maintenance and disposal.
  • Page 6: Technical Documentation

    • W@M Device Viewer: enter the serial number indicated on the nameplate (www.endress.com/deviceviewer). • The Endress+Hauser Operations App: enter the serial number indicated on the nameplate or scan the 2-D matrix code (QR code) provided on the nameplate. For a detailed list of the individual documents along with the documentation code (→...
  • Page 7: Registered Trademarks

    Registered trademarks ® ® Applicator , Nanomass Registered or registration-pending trademarks of the Endress+Hauser Group ® BOROFLOAT Registered trademark of Schott AG, Jena, Germany ® Microsoft Registered trademark of the Microsoft Corporation, Redmond, Washington, USA ®...
  • Page 8: Basic Safety Instructions

    Clarification of borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability as even minute changes in the temperature, concentration or level of contamination in the process can alter the corrosion resistance properties.
  • Page 9: Occupational Safety

    It meets general safety standards and legal requirements. It also complies with the EC directives listed in the device-specific EC Declaration of Conformity. Endress+Hauser confirms this by affixing the CE mark to the device. IT security We only provide a warranty if the measuring device is installed and used as described in the Operating Instructions.
  • Page 10: Product Description

    Product description Nanomass Gas Density Product description Product design 2 (3) A0026432 Fig. 1: Important components of the measuring device (device version with RS232 interface) Connection for 4-20mA analog outputs RS232 interface for digital signal transmission (optional) USB port for digital signal transmission and supply voltage (optional, not illustrated in graphic)
  • Page 11: Incoming Acceptance And Product

    • Enter the serial number indicated on the nameplate into the Endress+Hauser Operations App or scan the 2-D matrix code (QR code) on the nameplate with the Endress+Hauser Operations App: all the information about the measuring device is displayed.
  • Page 12 Incoming acceptance and product identification Nanomass Gas Density 4.2.1 Nameplate Order code: Ser. no.: Ext. ord. cd.: Dev.Rev.: p max: CRN no.: Tm/Ta: Density range: Suitable medium: Date: A0026851 Fig. 2: Example of a nameplate Place of manufacture Name of the measuring device Order code Serial number (ser.
  • Page 13 Nanomass Gas Density Incoming acceptance and product identification 4.2.2 Symbols on measuring device Symbol Meaning WARNING This symbol alerts you to a life-threatening situation. Failure to avoid this situation can result in serious or fatal injury. Protective ground connection A terminal which must be connected to ground prior to establishing any other connections.
  • Page 14: Storage, Transportation, Disposal Of

    Storage, transportation, disposal of packaging Nanomass Gas Density Storage, transportation, disposal of packaging Storage conditions Observe the following notes for storage: • Store in the original packaging to ensure protection from shock. • Do not remove protection caps mounted on connections.
  • Page 15: Installation

    Nanomass Gas Density Installation Installation Installation conditions No special measures such as supports are necessary. External forces are absorbed by the construction of the device. 6.1.1 Mounting position Mounting location The measuring device is usually installed in a bypass line. If flow rates are low, the device can also be installed in the main pipe.
  • Page 16 Installation Nanomass Gas Density Thermal insulation Due to a low thermal capacity of the medium, the medium temperature can be greatly affected by the ambient temperature through the supply line and the measuring device. The influence of the ambient temperature on the medium temperature can be reduced by insulating the supply line.
  • Page 17 Nanomass Gas Density Installation • Create necessary pressure drop with orifice plate in process pipe and with throttle (or flow monitor) downstream from the measuring device. ΔP < 0.1 bar (1.45 psi) P2 < P1 P1 < 20 bar (290 psi) A0026153 Fig.
  • Page 18: Mounting The Measuring Device

    Installation Nanomass Gas Density Swagelok pipe fitting The measuring device and the filter supplied are mounted in the pipe using a ¹⁄₄" Swagelok pipe fitting. Cut the pipe at a right-angle and deburr it. Push the pipe into the fitting as far as it will go.
  • Page 19: Post-Installation Check

    Nanomass Gas Density Installation Connect the filter to the pipe (→  15) and secure with a Swagelok pipe fitting (→  18). Mount the measuring device on a wall or ceiling using drill holes (→  16). Connect the measuring device to the pipe and secure with a Swagelok pipe fitting (→ ...
  • Page 20: Electrical Connection

    Electrical connection Nanomass Gas Density Electrical connection The measuring device does not have an internal circuit breaker. For this reason, assign the measuring device a switch or power-circuit breaker so that the power supply line can be easily disconnected from the mains.
  • Page 21 Nanomass Gas Density Electrical connection 7.1.3 Pin assignment of connector Connector for supply voltage Assignment Coding, Connector/ connector socket face DC 8 to 30 V Connector A0026825 Connector for signal transmission 4-20 mA connector Assignment Coding, Connector/ connector socket face...
  • Page 22: Connecting The Measuring Device

    Electrical connection Nanomass Gas Density Connecting the measuring device WARNING Improper connection can cause serious injury or death ‣ Have electrical connection work carried out by correspondingly trained specialists only. ‣ Observe applicable federal/national installation codes and regulations. ‣ Comply with local workplace safety regulations.
  • Page 23 Nanomass Gas Density Electrical connection 7.2.2 Connecting the signal cables Signals can be transmitted digitally via a USB port or RS232 interface or by analog communication via the passive 4-20 mA connection. Connect the 4-20 mA cable Strip 10 mm (0.4 in) of the cable and cable ends. In the case of stranded cables, also fit ferrules.
  • Page 24: Special Connection Instructions

    Electrical connection Nanomass Gas Density Special connection instructions 7.3.1 Connection examples Current output 4-20 mA 4...20 mA 4...20 mA A0026824 Fig. 8: Connection example for 4-20 mA current output Automation system with current input (e.g. PLC) Cable shield, comply with connecting cable requirements (→  20) Power supply 12-24 V Analog display unit: observe maximum load (→...
  • Page 25: Post-Connection Check

    Nanomass Gas Density Electrical connection Post-connection check Are cables and the device undamaged (visual inspection)?  Do the cables comply with the requirements (→  20)?  Do the cables have adequate strain relief?  Does the supply voltage match the specifications on the nameplate? ...
  • Page 26: Operating Options

    Operating options Nanomass Gas Density Operating options Overview of operating options The measuring device offers users the following operating options: • Local operation • "Nanomass Communication" operating tool (via USB port or RS232 interface) The operating language of the measuring device is English. Other language options are not supported.
  • Page 27: Operating Elements

    Nanomass Gas Density Operating options The number and display format of the measured values can be configured in the "User interface" → "Assign" menu. Parameterization views The user is in the operational display mode: press Enter to enter the parameterization mode.
  • Page 28: Access To The Measuring Device Via The "Nanomass Communication" Operating Tool

    Operating options Nanomass Gas Density 8.2.2 Functions Function Operation Display measured values Once the measuring device has been connected to the supply voltage and the measuring device is ready for operation, the measured values appear on the operational display. Enter parameterization The user is in the operational display mode.
  • Page 29 Nanomass Gas Density Operating options Computer Depends on the device interface: • USB port or • RS232 interface Disk drives (in the case of installation the "Nanomass Communication" operating tool via CD-ROM): • CD-ROM drive Computer software Recommended operating system •...
  • Page 30 Operating options Nanomass Gas Density First connection of the device If the device is connected the first time on the computer, it will incorrectly recognized. ‣ In case of first device connection install USB driver. Install USB driver: Depending on the version, connect the device to your computer via the USB- or the RS232-interface.
  • Page 31: User Interface

    Select the communication interface for communicating with the measuring device. The list displayed contains all the COM ports available. A "Refresh" function is also provided to update the list. Device type Device type Displays the type of device that is connected ("Nanomass Gas Density"). Endress+Hauser...
  • Page 32: Measured Value Display

    Operating options Nanomass Gas Density Device configuration and device functions Function area Description Parameter Parameter Access to the device' s operating menu structure; same as local display: • Configuration of parameters Real-time Real-time measuring data measuring data Access to current measured values: •...
  • Page 33 Nanomass Gas Density Operating options 8.3.7 Parameter Users can access the operating menu structure of the device in the "Parameter" function area. This menu structure can also be called via local operation. Screen layout A0027435 Fig. 11: Parameter Navigation area...
  • Page 34: System Integration

    System integration Nanomass Gas Density System integration Overview of device description files 9.1.1 Current version data for the device Firmware version • On the title page of the Operating instructions • On the nameplate (→  12) • In Nanomass Communication: function area "Device settings"...
  • Page 35 Nanomass Gas Density System integration 9.2.2 Data structure and commands The commands which can be used to read measured variables and device information out of the measuring device are listed below. The data are transmitted in packets. The data are divided into seven packets.
  • Page 36: Commissioning

    Commissioning Nanomass Gas Density Commissioning 10.1 Installation and function check Before commissioning the device, make sure that the post-installation and post-connection checks have been performed. • "Post-installation check" checklist →  19. • "Post-connection check" checklist →  25. 10.2 Switching on the measuring device After a successful function check, switch on the measuring device.
  • Page 37: Setting The Date And Time

    Nanomass Gas Density Commissioning Navigation Function area "Device settings" → "COM-Port → "Data retrieval delay" To access the "Device settings" function area, you must enter the access code 0074 in the "Access code for technician mode" field and then press "OK" to confirm.
  • Page 38 Commissioning Nanomass Gas Density 10.4.5 Configuring the current output Various parameters are used to configure the current output. In addition, there are differences depending on which current output is being configured. Navigation Menu "Measuring variables" → "System units" Menu "Output 4-20 mA" → "Output channel 1" and "Output channel 2"...
  • Page 39 Nanomass Gas Density Commissioning Parameter Procedure Selection/input Factory setting Pressure Select the unit for pressure. • bar absolute Country-specific: • bar gauge • kPa absolute Result • psi absolute • psi absolute The selected unit applies for: • psi gauge •...
  • Page 40 Commissioning Nanomass Gas Density Parameter Description Selection/input Factory setting Failsafe mode channel 2 Specify the output behavior of current • Min current Min current output 2 in the event of a device alarm. • Max current • Hold value Note •...
  • Page 41: Advanced Settings

    Nanomass Gas Density Commissioning Menu "User interface" → "Assign" → "Bottom line display" Parameter Description Selection/input Factory setting Bottom line display Select the measured value that is shown • Off Temperature on the second line of the local display. • Density •...
  • Page 42 Commissioning Nanomass Gas Density Navigation Menu "Measuring variables" → "Special units" Parameter overview with brief description Menu "Measuring variables" → "Special units" Parameter Description Selection/input Factory setting Text arb. density (4 char.) Enter the text string for the arbitrary XXXX (max. 4 characters) "___"...
  • Page 43: Simulation

    Nanomass Gas Density Commissioning Performing density adjustment with medium The measuring device must be filled with the medium to perform density adjustment. Furthermore, the exact density of the medium at the indicated temperature and pressure must be known. Call up the "Device settings" function area.
  • Page 44: Concentration Measurement" Application Package

    It is possible to subsequently activate the function. 10.7.2 Activation For activation, a conversion kit from Endress+Hauser is required. Among other things, this kit contains an activation code which must be entered via Nanomass Communication to activate the "Concentration Measurement" function.
  • Page 45 (e.g. methane or carbon dioxide) and display this information. 10.8.4 Determining the coefficients Your Endress+Hauser Sales Center can provide you with the coefficients that must be saved in the measuring device to calculate the concentration. The following information is needed to determine the coefficients: •...
  • Page 46 Commissioning Nanomass Gas Density Menu "Special function" → "Density function" Parameter Description Selection/input Factory setting Density type Select the density function for calculating • Flexible special density values or the percentage of components in binary gas mixtures. Concentration unit Select the unit for calculating the •...
  • Page 47: Operation

    Nanomass Gas Density Operation Operation 11.1 Changing the operating language The operating language of the measuring device is English. Other language options are not supported. 11.2 Configuring the local display Settings for the local display (→  40) 11.3 Reading measured values on the local display All the current measured values for every process variable can be read on the local display with the "Main values"...
  • Page 48: Accessing Real-Time Measuring Data

    Operation Nanomass Gas Density 11.4 Accessing real-time measuring data Real-time measuring data can be graphically displayed and recorded to a text file in the "Real-time measuring data" function area via Nanomass Communication. 11.4.1 Screen layout A0027437 Fig. 13: Real-time measuring data...
  • Page 49: Stopping Recording

    Nanomass Gas Density Operation Press the "Select" button.  A window opens where you can select where the file should be saved. Select the desired storage location. You can choose from the following options: – If you selected "Create new file" previously, select the desired storage location for the new file.
  • Page 50 Operation Nanomass Gas Density 11.4.8 Moving the Y-axis position of the graph Click the "Autoscale" button to switch off the autoscale function. Note: The autoscale function is switched off if the button is no longer blue and "OFF" is displayed.
  • Page 51: Retrieving Measuring Data

    Nanomass Gas Density Operation 11.5 Retrieving measuring data It is only possible to access the "Measuring data retrieval" function area, if the "Datalog function" application package was activated. Otherwise the function area cannot be accessed (→  43). Measuring data that are saved in the internal data memory can be retrieved in the "Measuring data retrieval"...
  • Page 52: Adapting The Measuring Device To The Process

    Operation Nanomass Gas Density Select the file where the real-time measuring data will be saved. You can choose from the following options: – Select "Create new file" to save measuring data in a new file. – Select "Extend existing file" to insert measuring data into an existing file.
  • Page 53: Updating The Firmware

    Nanomass Gas Density Operation 11.7 Updating the firmware The firmware must be updated using the "Nanomass Communication" operating tool (digitally via a USB or RS232 interface). Navigation Function area "Device settings" → "Service mode" To access the "Device settings" function area, you must enter the access code 0074 in the "Access code for technician mode"...
  • Page 54: Diagnostics And Troubleshooting

    Diagnostics and troubleshooting Nanomass Gas Density Diagnostics and troubleshooting 12.1 General troubleshooting For local display Problem Possible causes Remedy Local display dark and no Supply voltage via supply voltage cable: output signals. LEDs on the The supply voltage does not match Apply the correct supply voltage (→...
  • Page 55: Diagnostic Information Via Light Emitting Diodes

    Nanomass Gas Density Diagnostics and troubleshooting For access via the "Nanomass Communication" operating tool Error Possible causes Solution Not possible to establish connection Cable is too long. Check the cable length and correct if to the "Nanomass Communication" necessary. operating tool.
  • Page 56: Diagnostic Information In The Operating Tool

    Diagnostics and troubleshooting Nanomass Gas Density 12.4 Diagnostic information in the operating tool Any faults detected by the measuring device are indicated in the operating tool by red LEDs under "Device state" in the "Device settings" function area. Diagnostics message...
  • Page 57: Performing Density Adjustment With Air

    Nanomass Gas Density Diagnostics and troubleshooting 12.6 Performing density adjustment with air It is possible to perform density adjustment with air for diagnostic purposes. The density adjustment must be performed using the "Nanomass Communication" operating tool. Navigation Function area "Device settings" → "Calibrate density" → "Density calibration using air"...
  • Page 58 Diagnostics and troubleshooting Nanomass Gas Density Parameter Description Display Factory setting Density lower bound Displays the lower density Positive integer limit. Tag name Displays the tag name. Character string "Your info here" consisting of letters and numbers Endress+Hauser...
  • Page 59: Maintenance

    13.2 Measuring and test equipment Endress+Hauser offers a wide variety of measuring and test equipment, such as W@M or device tests. Your Endress+Hauser Sales Center can provide detailed information on the services.
  • Page 60: Endress+Hauser Services

    Maintenance Nanomass Gas Density 13.3 Endress+Hauser services Endress+Hauser offers a wide variety of services for maintenance such as recalibration, maintenance service or device tests. Your Endress+Hauser Sales Center can provide detailed information on the services. Endress+Hauser...
  • Page 61: Repair

    • Spare parts are grouped into logical kits with the associated Installation Instructions. • Repairs are carried out by Endress+Hauser Service or by appropriately trained customers. • A certified measured device can only be converted to another certified device by Endress+Hauser Service or at the factory.
  • Page 62: Disposal

    Repair Nanomass Gas Density 14.5 Disposal 14.5.1 Removing the measuring device Switch off the measuring device. Warning! Danger to persons from process conditions. Beware of hazardous process conditions such as pressure in the measuring device, high temperatures or aggressive fluids. Carry out the mounting and connection procedure described in the "Mounting the measuring device"...
  • Page 63: Accessories

    Various accessories are available for the measuring device, and can be ordered with the measuring device or at a later stage from Endress+Hauser. Detailed information on the appropriate order code is available from your local Endress+Hauser Sales Center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 64: System Components

    Accessories Nanomass Gas Density 15.3 System components Accessories Description Memograph M graphic The Memograph M graphic display recorder provides information on all relevant display recorder measured variables. Measured values are recorded correctly, limit values are monitored and measuring points analyzed. The data are stored in the 256 MB internal memory and also on a SD card or USB stick.
  • Page 65: Technical Data

    Nanomass Gas Density Technical data Technical data 16.1 Application The measuring device is suitable for density and concentration measurement of gases only. Depending on the version ordered, the measuring device can also measure potentially explosive, flammable, poisonous and oxidizing media.
  • Page 66: Power Supply

    Technical data Nanomass Gas Density Resolution 8 μA Damping Configurable: 0 to 120 s Assignable measured variable • Density • Reference density • Concentration • Temperature • Pressure Signal on alarm Breakdown information is displayed as follows. Current output Failure mode...
  • Page 67: Performance Characteristics

    Nanomass Gas Density Technical data Current consumption Maximum switch-on current 1 A (< 0.125 ms) Power supply failure • The configuration and recorded data are retained in the data memory. • Internal real-time clock is battery-backed and continues to run. The operating life of the real-time battery is 10 years.
  • Page 68: Installation

    Technical data Nanomass Gas Density Repeatability Base repeatability under reference operating conditions Density (gases) ±0.05 kg/m (±0.00005 g/cm Temperature ±0.25 °C Pressure ±0.02 bar Response time <200 ms Influence of medium The medium temperature does not affect the measuring accuracy. The effects of temperature temperature are compensated for by measuring the temperature in the micro-channel.
  • Page 69: Process

    • Propene (C • Propane (C • Butane (C • LPG (supplied as gas) If you are using gases other than those indicated above, please contact your Endress+Hauser Sales Center at: www.endress.com/worldwide. Medium temperature –20 to +60 °C (–4 to +140 °F)
  • Page 70 Technical data Nanomass Gas Density Pressure-temperature The following pressure/temperature diagram refers to the entire measuring device and not ratings just the process connection. [psi] [bar] [°C] [°F] A0026823-DE Secondary containment The sensor housing is filled with dry nitrogen on delivery and protects the electronics and pressure rating mechanics inside.
  • Page 71: 16.10 Mechanical Construction

    Nanomass Gas Density Technical data 16.10 Mechanical construction Design, dimensions 157.5 (6.2) ~165 (~6.5) 55.75 (2.2) 23 (0.9) 142 (5.6) A0026010 5.5 (0.2) 157.5 (6.2) 143.5 (5.65) 7 (0.26) 7 (0.26) 16 (0.63) 4 x max. SW 10 A0026011 Weight Ca.
  • Page 72: 16.11 Operability

    Technical data Nanomass Gas Density Materials Housing • Aluminum, powder-coated • Window material: polycarbonate Connections 4-20 mA connection • Socket: die-cast zinc, nickel-plated • Contact housing: polyamide • Contacts: brass, gold-plated RS232 interface • Socket: die-cast zinc, nickel-plated • Contact housing: polyamide •...
  • Page 73: 16.12 Certificates And Approvals

    16.12 Certificates and approvals CE mark The measuring system meets the legal requirements of the EC Directives. Endress+Hauser confirms that the device has been successfully tested by applying the CE mark. C-Tick symbol The measuring system complies with the EMC requirements of the "Australian Communications and Media Authority (ACMA)".
  • Page 74: 16.14 Accessories

    • W@M Device Viewer: enter the serial number indicated on the nameplate (www.endress.com/deviceviewer). • The Endress+Hauser Operations App: enter the serial number indicated on the nameplate or scan the 2-D matrix code (QR code) provided on the nameplate. Standard documentation...
  • Page 75: Appendix

    Nanomass Gas Density Appendix Appendix 17.1 Overview of the device parameters The following table provides an overview of the menu structure and the parameters in the individual menus. The page reference indicates where the associated description of the submenu or parameter can be found. Parameters marked with an asterisk (*) are only visible on the local display.
  • Page 76: Menu "Measuring Variables" → "Main Values

    Appendix Nanomass Gas Density Simulation Activate simulation Current value Basic function System parameter →  84 Time constant input Special function Density function →  84 Density type: Fixed Reference pressure Reference temperature Density type: Flexible Concentration unit Concentration names...
  • Page 77: Menu "Measuring Variables" → "System Units

    Nanomass Gas Density Appendix Oscillation frequency Description Displays the oscillation frequency of the measuring channel. 17.3 Menu "Measuring variables" → "System units" Density Description Use this function to select the unit for the density. Options • g/cm³ • g/cc • kg/L •...
  • Page 78: Menu "Measuring Variables" → "Special Units

    Appendix Nanomass Gas Density Pressure Description Use this function to select the unit for the medium pressure during density measurement. Options • bar absolute • bar gauge • psi absolute • psi gauge • kPa absolute • kPa gauge Factory setting Country-specific: •...
  • Page 79: Menu "User Interface" → "Display

    Nanomass Gas Density Appendix Factor arbitrary concentration Prerequisite The optional "Concentration Measurement" application package must be installed. Description Use this function to enter a quantity factor for the arbitrary concentration unit. User entry 6-digit floating point number and a 1-digit exponent Factory setting 1.00000E+0...
  • Page 80 Appendix Nanomass Gas Density Top line display multiplex Description Use this function to assign a process variable to the first line of the local display when the display changes. Options • Off • Density • Ref. density • Concentration • Temperature •...
  • Page 81: Menu "Output 4-20 Ma" → "Output Channel 1

    Nanomass Gas Density Appendix 17.7 Menu "Output 4-20 mA" → "Output channel 1" Assign Description Use this function to assign a process variable for current output 1. Options • Off • Density • Ref. density • Concentration • Temperature • Pressure...
  • Page 82: Menu "Output 4-20 Ma" → "Output Channel 2

    Appendix Nanomass Gas Density 17.8 Menu "Output 4-20 mA" → "Output channel 2" Assign Description Use this function to assign a process variable for current output 2. Options • Off • Density • Ref. density • Concentration • Temperature • Pressure...
  • Page 83: Menu "Output 4-20 Ma

    Nanomass Gas Density Appendix 17.9 Menu "Output 4-20 mA" → "Response channel 1/2" Failsafe mode channel 1 Description Use this function to specify behavior in the event of a device alarm. Options • Min current • Max current • Hold value •...
  • Page 84: 17.11 Menu "Basic Function" → "System Parameter

    Appendix Nanomass Gas Density Current value Prerequisite The "On" option is selected in the "Activate simulation" parameter. Description Use this function to enter a current value for the simulation. User entry 2.00 to 23.00 mA Factory setting 2.00 mA Additional information If simulation is enabled ("On"...
  • Page 85 Nanomass Gas Density Appendix Reference temperature Prerequisite This function is only available if the "Fixed" option was selected in the "Density type" function. Description Use this function to enter a reference temperature for calculating the reference density. User entry Decimal number with sign Factory setting 20 °C...
  • Page 86: 17.13 Menu "Supervision

    Appendix Nanomass Gas Density Concentration offset Prerequisite This function is only available if the "Flexible" option was selected in the "Density type" function. Description Use this function to enter an offset value for the calculated concentration. User entry Decimal number with sign -5.000 to 5.000...
  • Page 87 Nanomass Gas Density Appendix Factory reset Description Use this function to select whether all the device settings are to be reset to factory defaults. Options Local operation: • Yes • No Factory setting Local operation: No Additional information Local operation: If "Yes" is selected, all the device settings are reset to the factory setting and the measuring device is restarted.
  • Page 88: Index

    Nanomass Gas Density Index Device Parameters....... . 75 Overview........75 Accessories .
  • Page 89 Nanomass Gas Density Input ......... . 65 Inspection check Nameplate .
  • Page 90 Nanomass Gas Density Requirements for personnel ......8 Response time ........68 Use of the measuring device Return.
  • Page 91 Nanomass Gas Density Endress+Hauser...