Programs; Using Mask-Fit - ResMed Stellar 150 Clinical Manual

Invasive and noninvasive ventilator
Hide thumbs Also See for Stellar 150:
Table of Contents

Advertisement

Programs

Using mask-fit

The patient can use mask-fit to help you fit the mask properly. This feature delivers constant
treatment pressure for a three-minute period, prior to starting treatment, during which the
patient can check and adjust the mask-fit to minimize leaks. The mask-fit pressure is the set
CPAP or EPAP pressure or 10 cm H
1
Instruct the patient to lie down in bed or in their typical usage position and to put the mask on
according to the mask user guide.
2
Hold down
3
Adjust the mask, mask cushion and headgear until the patient has a good mask-fit.
After three minutes at the mask-fit pressure, the prescribed treatment mode and pressures will
commence. Mask-fit can be stopped at any time by pressing
Notes:
Press
for at least three seconds during mask-fit to start treatment immediately. The
Treatment screen will be displayed.
The mask-fitting feature is disabled when the mask type Trach is selected.
Availability of this feature is country dependent.
26
Ramp time
Up to two sets of therapy settings, ie, Programs, can be
Program
saved for convenient retrieval. They are stored while in the
Clinical Settings menu. If both, ie, dual, is selected, the
patient can choose the program to use on the Treatment
screen. If only a single program is selected, the option does
not display.
Single or dual programs can be set in the Setup menu,
Options, Configuration menu.
The Program names can be personalized using ResScan (eg,
Day and Night).
O, whichever is greater.
2
for at least three seconds until pressure delivery starts.
IPAP
EPAP
Start pressure
0
Ramp time
.
Time

Hide quick links: