Adjusting Device Settings; Programs - ResMed Astral 100 Series User Manual

Adult and pediatric ventilator
Table of Contents

Advertisement

Device setting
Description
Date
Allows setting of the day, month and year of the current date.
Time
Allows setting of the hours and minutes of the current time.
Language
Sets the current language of the device selected from the list of available languages.
Pressure unit
Specifies displayed units for all pressure data and settings as cmH
Note: The reference unit for all accuracy and monitoring claims is hPa. The conversion factor between
units in accordance with industry practice is one.
Height unit
Specifies displayed units for the patient's height as inches or cm.

Adjusting device settings

Access adjustable device settings from the Setup menu and select Device Config.
The current active selections are highlighted in orange.
To change settings, simply select another of the available options. The revised setting is highlighted in
orange.

Programs

Programs on the Astral device can be configured by your clinician to provide you with alternate treatment
options. For example, a clinician can set up programs for sleeping versus daytime use, or for use during
exercise or physiotherapy. Programs allow for different circuit, ventilation and alarm settings.
The Astral device comes with one standard active program. Your clinician can configure up to three
additional programs (if available).
If any additional programs have been set up by your physician, they can be selected for use from the
Patient home screen. You can change between programs while the Astral device is delivering ventilation.
Changing between programs will cause ventilation and alarm settings to change, as configured by your
clinician.
Using the Astral device
O, mBar or hPa.