Recommended procedure for message 82, 94 and 111
The messages 82, 94 and 111 can occur when the radiating unit
·
·
·
·
To ensure normal operation, GE Sensing & Inspection Technologies recommends conditioning
the radiating unit in power mode. To achieve this, the following operating parameters are
required:
After successful operation in power mode as described above, normal operation can be
resumed.
Note:
Copyright by GE Sensing & Inspection Technologies GmbH
62
Operating Manual ERESCO MF4 Type Series
is only operated occasionally at the maximum high voltage.
has not been operated for a longer period or is only operated occasionally.
is operated for short operating cycles.
is operated in extremely high or extremely low ambient temperatures.
ERESCO 160 MF4-R
ERESCO 200 MF4-R
ERESCO 42 MF4
ERESCO 32 MF4-C
ERESCO 65 MF4
ERESCO 52 MF4-CL
ERESCO 280 MF4-R
ERESCO 300 MF4-R
If after execution of the said procedure the radiating unit cannot be
operated at the desired high voltage, the power mode procedure can be
repeated up to three times. If this does not result in correct operation of
the radiating unit, please contact the Customer Service Department of
GE Sensing & Inspection Technologies.
60 kV
10 mA
60 kV
10 mA
90 kV
10 mA
60 kV
10 mA
150 kV
6 mA
100 kV
6 mA
100 kV
3.4 mA
150 kV
6 mA
Bogenstrasse 41 D - 22926 Ahrensburg
10 min
10 min
10 min
10 min
10 min
10 min
10 min
10 min
08/24/2016