Invacare Birdie User Manual page 19

Portable patient lift
Hide thumbs Also See for Birdie:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 3
1.
Lower the boom by pressing and holding button A at the front
of the control unit.
2.
Stop lowering the boom by releasing the button.
Emergency Release with Jumbo Care control unit
C
A
Insert a pen into the hole labeled Emergency Up A or Emergency
1.
Down B on the control unit C.
Activating an emergency release manually
(Not available on Birdie™Compact)
For the case of partial or total power failure, or if the battery
runs down while using the lift, Birdie™is equipped with a manual
emergency release system located at the bottom of the actuator.
It is recommended that the primary emergency release be
used. The secondary (manual) emergency release is only a
back-up to the primary emergency release.
1575480-D
1.
B
The manual emergency release system will only operate with a
patient in the lift. It can be adjusted according to the patient's weight
as described below. The weight is preset to 75 kg.
Adjusting the lowering speed for manual emergency
lowering
1.
Locate the screw on top of the red emergency grip A.
2.
Loosen the screw to increase the speed.
3.
Tighten the screw to decrease the speed.
Pull up on the emergency
grip A and push down on
the boom at the same time.
Usage