Updating Driving Program - Invacare Colibri Service Manual

Hide thumbs Also See for Colibri:
Table of Contents

Advertisement

3.
Pull shroud (1) on operating console upwards out of
the fixing.
4.
Remove screws (1) from operating console.
5.
Remove all plugs from operating console.
6. Remove operating console.
7.
Loosen and remove screws (1) including drive lever.
8.
Remove screws C on drive lever mounting B and
remove mounting.
1577883-B
Fig. 4-7
Fig. 4-8
Fig. 4-9
Fig. 4-10
Fig. 4-11
9. Disconnect potentiometer plug A from operating
console.
10.
Remove screws A on potentiometer.
11. Replace potentiometer.
12. Reconnect potentiometer plug A with operating
console.
13.
Connect potentiometer A and drive lever mounting B.
Tighten screw C.
14. Install parts in reverse order.
15. Test all functions (trial run).

4.5.4 Updating driving program

The driving programs for electric wheelchairs are
continually updated and improved by Invacare®. For this
reason, you should check whether the version number
is still up to date each time a wheelchair comes in for
repairs, and also during regular inspections.
If a newer version is available, the driving program must be
updated. The procedure for updating the driving program
is described in the user manual of the Wizard software.
The electronic system is supplied with a standard
drive program. If the driving program has been
customised, you have to perform this customisation
again, after installing the new driving program.
Dynamic® Wizard software
User manual for the Wizard software
For further information on other requirements
- such as the minimum system configuration of
the PC to be used for programming, necessary
programming cables - see the user manual
of the Wizard software. You find the latest
version of the user manual in the download
area on http://www.dynamiccontrols.com/.
Fig. 4-12
Fig. 4-13
Service
13