Mfcmu - Multi Frequency Cmu - Agilent Technologies B1500A User Manual

Semiconductor device analyzer
Hide thumbs Also See for B1500A:
Table of Contents

Advertisement

MFCMU - Multi Frequency CMU

This section describes typical specification of the multi frequency capacitance
measurement unit (CMU) for Agilent B1500A. The CMU performs the impedance
measurement and returns the specified measurement data such as Cp-G. Only one
module can be installed in one mainframe.
Measurement parameters: See Table 2-13.
AC signal, output frequency:
1 kHz to 5 MHz
Setting resolution: 1 mHz (1 kHz to), 10 mHz (10 kHz to), 0.1 Hz (100 kHz to),
or 1 Hz (1 MHz to 5 MHz)
AC signal, output level:
10 mVrms to 250 mVrms, 1 mV resolution
DC bias:
0 to ± 25 V, using MFCMU internal DC bias
0 to ± 100 V, using SMU and SCUU (SMU CMU unify unit)
Measurement range:
For the fixed ranging mode, measurement range (impedance range) must be
specified to perform measurement. Table 2-14 shows the available measurement
ranges and the corresponding impedance value. And Figure 2-6 shows the
calculation example of the impedance vs frequency characteristics of a
capacitive load. Use these information to decide the measurement range.
Impedance Z is calculated by the following formula.
Z = 1 / (2 f C)
where f is frequency (Hz) and C is capacitance (F).
Agilent B1500A User's Guide, Edition 10
Introduction
Measurement Units
2- 29