4
Performance Tests and Calibration
80
Downloaded from
www.Manualslib.com

Optional AC Current Performance Verification Test

Configuration: AC Current (CONFigure:CURRent:AC)
1 Select the AC Current function.
2 Select each range in the order shown below. Provide the indicated input
voltage and frequency. Compare measurement results to the appropriate
test limits shown in the table. (Be certain to allow for appropriate
source settling.)
Table 12
Optional AC Current Performance Verification Test
Current
Input Frequency Function
10mA
20Hz
10mA
45Hz
10mA
1kHz
10mA
10kHz
1A
1kHz
100mA
1kHz
10mA
1kHz
[1] Select 5½- digit measurement resolution
manuals search engine
Error from Nominal
[1]
Range
1 year
AC Current
10mA
± 0.16mA
10mA
± 0.16mA
10mA
± 60µA
10mA
± 0.22mA
1A
± 6mA
1A
± 1.5mA
1A
± 1.05mA