White Calibration

When using the instrument for long periods of time, the displayed value may change depending on
changes in the environment. Therefore, in order to achieve accurate measurements, we recommend
that white calibration is done regularly using the white calibration plate. White calibration should also
be done before measuring after a long time has passed since the instrument was last used.
White calibration should be done under the same temperature conditions as measurement.
1
• The white calibration screen appears.
[Operating Procedure]
1
Press the Calibration key
No white calibration
data has been set.
• The cursor moves automatically to the next position when setting data using the numeric pad.
• Use the
• Go to step 3, if white calibration data has already been set.
[Example]
Set the calibration data listed on the back of the white calibration plate cover
2
• Illuminants C
using the
• Illuminants D
86
Color Space key
2
4
while in the measurement screen.
to move the cursor to make corrections.
Y=93.5 x=0.3114 y=0.3190
key and the numeric pad.
Y=93.5 x=0.3140 y=0.3318
65
(4)
3
White calibration data
Ready lamp
White calibration plate
has been set.