Acting in case of an alarm
MAGNETOM Skyra
fit
| syngo MR E11
Print No. M6-04401.621.02.02.02
Gradient malfunction is detected.
An alarm message appears at the syngo Acquisition Workplace.
Dialog box at the syngo Acquisition Workplace
(1) Scanner hardware error
Automatic shutdown! Evacuate patient immediately from the
examination room! To prevent further damage, the system will
be switched to Standby within 1 minute! Call Siemens Service
immediately.
1
Immediately move the patient out of the magnet bore by pressing
the Home Position button.
Within the next 60 seconds, the system will be switched to
Standby automatically. Then, the patient table motors can not be
operated any longer.
2
Call Siemens Service.
MR system components
3
81