After the Magnet Stop switch has been pressed, an alarm is triggered
at the alarm box. The WARNING LED will light up and an alarm signal
will sound.
As a rule, Siemens Service must be called following a quench.
The magnet must only be put back into operation by Siemens
Service personnel.
 WARNING
Formation of droplets due to condensation during quenching!
Personal injury
Risk of fire
Do not touch the exhaust line.
Do not stand under the exhaust line.
Avoid open flames and do not smoke.
 WARNING
System indicates Magnet Stop error!
Hazardous conditions because the magnet cannot be
quenched in case of emergency
Immediately remove the patient from the magnet.
Restrict the access to the examination room.
Notify Siemens Service.