Safety
System-related artifacts/
imaging errors
Stripe artifacts
Incorrect slice positioning
34
Causes: Artifacts and imaging errors are listed according to their
source for error:
System-related artifacts/imaging errors
Patient-related artifacts/imaging errors
User-related artifacts/imaging errors, see:
manual
User-related and patient-related artifacts/imaging errors can be
largely avoided through patient instructions and proper conduct of
patient and personnel.
The MR image may show system-related artifacts/imaging errors
despite careful preparation.
If the same artifact/imaging error appears repeatedly, document
and submit it to Siemens Service.
 CAUTION
RF-signal interference caused by incompatible accessories e.g.
patient monitoring devices!
Streaks and bright spots in the MR image
Use only accessories tested and approved for the MR system.
Keep the door to the examination room closed.
Vary the bandwidth of the MR sequence.
Whenever possible, use local coils for the MR examination.
 CAUTION
Phasing of MR signal is not set correctly!
Structure is shown in the wrong position
Repeat the measurement for the structure in question by
using a second orthogonal slice and check whether the
position of the structure is reproducible nor not.