Variations in brightness
Variations in signal and
MAGNETOM Skyra
fit
| syngo MR E11
Print No. M6-04401.621.02.02.02
CAUTION
Local variation in the sensitivity of local coils!
Continuous fluctuations in MR image brightness
◆
Whenever possible, use a local coil with transmit
characteristics that are more suitable for the field of view
desired.
◆
Use the normalization filter.
CAUTION
Static and/or stationary brightness errors on the LCD monitor
due to aging!
Incorrect diagnosis
◆
Scroll through the images to ensure that the MR image does
not show differences in brightness, spots, or cloudiness and
check bright objects for afterglow.
◆
Keep your eyes always in central monitor position with a
vertical view angle towards the screen surface for best image
quality.
CAUTION
contrast
Inhomogeneous RF field!
Asymmetry of contrast in the MR image
◆
Whenever possible, use a local coil with transmit
characteristics that are more suitable for the desired field of
view.
Safety
2
35