Troubleshooting: Problems
during undocking
Undocking in case of
emergency
MAGNETOM Skyra
fit
| syngo MR E11
Print No. M6-04401.621.02.02.02
MR system components
Table moves downwards despite you want to lift the table
In certain cases (for example, after docking), the table must find
the reference position and moves slightly down. This is normal
behavior.
Table does not move horizontally
Check that the side rails are fully stored in the carrying frame - oth-
erwise no table movement is possible.
Pedal cannot be pressed and the table is not pushed away from
the MR System
Set the lateral pedal to the middle position to deactivate the brake
and make sure that the tabletop is completely in Home position.
If undocking via the pedals does not work, the table can be undocked
by using the emergency undocking handle at the foot end (between
the pedals).
 WARNING
Wrong tabletop position for emergency undocking!
Uncontrolled table movement, tabletop crashes
Use the emergency undocking handle only when the tabletop
is in the outermost position (Home position).