MR system components
Stopping table movement
Rescuing the patient in an
emergency
Rescuing the patient manually
62
1
Press the Table Stop button to stop the table movement.
2
To reset the table stop, turn the Table Stop button clockwise until
it releases mechanically. Then simultaneously press the Table Up/
Inward and the Table Down/Outward button fully.
In case of accidents, e.g. patient emergency situation (e.g. heart
attack), the tabletop and patient must be moved out of the magnet
bore.
The fastest method for moving the tabletop out of the magnet
bore is to press the Home Position button. Select this method
whenever the power supply and/or motorized drive are intact.
1
Press the Home Position button.
The tabletop moves completely out of the magnet.
2
Rescue the patient.
In case of power failure and/or defective motorized drive, pull the
tabletop manually out of the magnet bore.
MR System | Operator Manual