FUNCTIONS ON THE UPPER SIDE
1. Suspension pin for connecting the hoist to the
hoist trolley.
2. Finger screw for dismounting the top section.

3.2 Operation

Hand control and indicator light
Robin
is activated when the hand control buttons are pressed. The blue indicator light illuminates when Robin
TM
is in use, and turns off automatically after approx. 30 seconds.
The hoist is lifted and lowered by operating the hand control. The control buttons on the underside of the hoist
may also be used for this purpose. The control buttons work independently of the hand control.
Note: The buttons must be depressed in the middle in order to ensure correct function, which is best obtained as
shown in the pictures.
The hoist will automatically run at high speed upwards as well as downwards, when it is not loaded.
The straps must be straight to ensure that they feeds into the apertures correctly. The straps are held taught by
the weight of the hooks, therefore the lifting straps should be freely suspended when the unloaded hoist is running
upwards and downwards.
When the hoist is loaded with a weight of more than 10 kg, the speed will automatically be lowered to a safe level
when moving in both directions.
Robin
has a built-in soft start, ensuring a more comfortable lifting and lowering movement of the user.
TM
Transport/running in the rail system
The empty hoist can be moved along the rail system by pulling the two lifting straps. The hand control must not be
used to pull the hoist along. During a transfer, the carer should gently push or pull the client sitting in the sling.
12
1
2
Operating forces of buttons:
Max. 5 N
Up
Down
TM