9.2.20 Calibration Signal Settings
Access Level: Engineer
Note: These settings refer to the range of signal values from the sensor over which calibration will be permitted, and are
used to avoid incorrect calibration e.g. due to incorrect application of gas.
Parameter Name
Maximum Zero
Offset
Minimum Zero
Offset
Maximum Span
Signal
Minimum Span
Signal
9.2.21 Calibration Span Gas Settings
Access Level: Engineer
These settings refer to the range of gas concentration values that can be used for span calibration, and are used to ensure
that the sensor is calibrated at an appropriate level.
The Maximum Span Gas Concentration and Minimum Span Gas Concentration are the maximum and minimum values
that can be entered into the Span Gas Concentration field during calibration.
The Default Span Gas Concentration is the value initially displayed in the Span Gas Concentration field. This may be
changed by the user during calibration to match the actual gas concentration used.
Parameter Name
Maximum Span Gas
Concentration
Minimum Span Gas
Concentration
Default Span Gas
Concentration
Note: MN and NN depend on the sensor type selected.
Part. No. 2400M2501_6
CONFIGURATION
Default
Value Range
Setting
5 mA
1 to 15 mA
3 mA
1 to 8 mA
20.0 mA
8.0 to 24.0 mA
5 mA
0.0 to 12.0 mA
Default Setting
Gas Range Max
Gas Range Min + MN * (Gas
Range Max - Gas Range Min)
Gas Range Max - NN * (Gas
Range Max - Gas Range Min)
Comment
During calibration, the 'Zero Signal Too High' warning is
set if the zero signal exceeds the Maximum Zero Offset
During calibration, the 'Zero Signal Too Low' warning is
set if the zero signal is below the Minimum Zero Offset
During calibration, the 'Calibration Signal Too High'
warning will be set if the span signal exceeds the
Maximum Span Signal.
During calibration, the 'Calibration Signal Too Low'
warning will be set if the span signal is below the Minimum
Span Signal
Value Range
Minimum
Gas Range Max- 0.20 * (Gas
Range Max-Gas Range Min)
Gas Range Min + 0.15 * (Gas
Range Max - Gas Range Min)
Gas Range Max - 0.70 * (Gas
Range Max-Gas Range Min)
73
Maximum
Gas Range Max
Gas Range Max - 0.20 * (Gas
Range Max - Gas Range Min)
Gas Range Max
Touchpoint Pro
Technical Handbook