C A U T I O N
1 Click File > Exit to close the 86100C application.
2 Click Start to view the Windows Start Menu.
3 Click My Computer and select the USER (D:) drive.
4 Click File > Format > Start and then OK. Do not change the default setup.
5 When the Format Complete dialog box appears, click OK.
6 Click Close.
To restore the 86100C Drive-D
Drive D on the 86100C hard disk contains all user files including data files. It
also includes backup copies of the factory calibration and application license
files that are automatically created from Drive C by the instrument. To restore
drive D back to factory defaults, you must reformat the partition using the
Windows XP Format utility. Formatting will erase all data on drive D.
Performing a Drive C Recovery immediately after formatting the D partition (without restarting
the instrument) results in the loss of factory calibration and any application license files.
Always turn off and then restart the instrument after performing the Drive-D Restoration
procedure.
Figure 2-17. Windows Start Menu
Leave the Format Options selections unchecked.