Running a Method or a Sequence from the Software Keypad

Loading, Storing, and Running Sequences from the Software Keypad

To load a sequence

To store a sequence

To start running a sequence

50
A sequence can specify up to five subsequences to be run,
as well as post- run sequences, if defined. Each sequence is
stored as a number (from 1 to 9).
Press [Load][Seq].
1
Enter the number of the sequence to be loaded.
2
Press [On/Yes] to load the sequence or [Off/No] to cancel
3
the load.
An error message is displayed if the specified sequence
number has not been stored.
Press [Store][Seq].
1
Enter the number of the sequence to be stored.
2
Press [On/Yes] to store the sequence or [Off/No] to cancel
3
the store.
1
Load the sequence. (See
Press [Seq Control].
2
3
Verify the status of the sequence:
Running—the sequence is running
Ready/wait—the instrument is not ready (due to oven
temperature, equilibration times, etc.)
Paused—the sequence is paused
Stopped—proceed to
Aborted—the sequence stopped without waiting for the
run to finish (See
No sequence—the sequence is off or not defined
Scroll to the Start sequence line and press [Enter] to change
4
the status to Running.
"To load a
sequence".)
step 4
"Aborting a
sequence".)