Entering a Method
Storing and Loading Methods
Storing a method
CAU TI O N
102
Store Method places the method in active memory into one of
four storage locations for future use. To store the method
currently in use:
1 Press Menu. This displays the following screen:
Load Method
Store Method
Chain Methods
Pressures
2 Use the cursor keys to highlight Store Method.
3 Enter the desired storage location {1–4} on the numeric
keypad followed by Enter.
The Headspace Sampler stores the current method in the
specified memory location. A stored method includes the
following parameters:
• Zone Temps
• Event Times
• Vial Parameters
• The advanced functions for Oven Stabilization Time
• The advanced functions for Pressure Units
• The advanced functions for Multiple Headspace Extraction
Storing overwrites any method already in that location. There is no
warning message that an overwrite is taking place.
DIAGNOSTIC
MAIN MENU
READY
↓
User Information