Contents 7820A GC Installation Guide Tools and Parts Required for Installation 7820A GC Installation Place and Prep the GC 7 Connect the checkout gases 10 Connect the cabling 7693A Automatic Liquid Sampler installation 18 Prepare for checkout 20 Set the PC IP address 21...
Page 4
Cable Diagrams Analog cable, general use Remote start/stop cable For More Information 7820A GC Installation Manual...
The installation requires the use of the Agilent GC and GC/MS User Manuals & Tools DVD that ships with your 7820A GC along with the available documentation for your GC (and Automatic Liquid Sampler, if purchased).
7820A GC Installation Guide Tools and Parts Required for Installation Agilent provides all specialized tools needed for installation. In addition, Agilent supplies several parts and plumbing assemblies to make installation easier. Find these tools and parts in the GC shipping container: •...
Place and Prep the GC Verify the site has been prepared for the GC. • Refer to the 7820A GC Site Preparation Guide on the GC and GC/MS User Manuals & Tools DVD. • Make sure the site meets the requirements in the guide, including power, and that clean gases and connection hardware are available.
Page 8
7820A GC Installation Guide Unpack the GC. Locate the tools, consumables, and the hardware user information & utility DVD. See Figure Big universal trap GC and GC/MS User Manuals & Tools DVD Agilent Technologies GC and GC/MSD Hardware User Information & Utilities Version B.01.01...
Page 9
7820A GC Installation Guide Place the GC on the bench and remove the detector caps under the detector cover. On the back panel, remove the caps. Check the power requirements on the GC back panel. Make sure the available power meets the requirements.
7820A GC Installation Guide Connect the checkout gases Install the Big Universal Trap into the carrier gas supply. See Figure • Install near the GC. • Follow the instructions included with the trap. • Use the nuts and ferrules supplied with the trap.
Page 11
7820A GC Installation Guide Optional shutoff valve To GC Indicating oxygen trap (recommended) Big Universal Trap Carrier gas supply To gas To GC supply Big Universal Trap Figure 2 Trap installation Select the pre-swaged gas tubing for your GC. See Figure •...
Page 12
7820A GC Installation Guide G4331-60003 Use to connect a single supply gas as: • Inlet carrier gas • FID, FPD, or NPD makeup gas G4332-60004 FID, FPD, and NPD air and hydrogen must be Use to connect a single supply gas as: supplied through separate tubing.
Page 13
7820A GC Installation Guide Install the tubing and connect to gas supply. See Figure Figure 5, and Figure Figure 4 Connecting Swagelok tubing 7820A GC Installation Manual...
Page 15
7820A GC Installation Guide uECD Helium carrier Nitrogen anode purge and makeup Figure 6 Proper plumbing configuration for instrument checkout (uECD) Set the gas source pressures. Recommended Maximum Helium 400 kPa (60 psi) 690 kPa (100 psi) Hydrogen 400 kPa (60 psi)
Page 16
7820A GC Installation Guide Check for external leaks. Maintain pressure for 5 minutes. Turn off gas supplies. Wait 10 minutes. The pressure in the supply lines should remain constant. If not, check for and fix any leaks. Restore supply pressures when leak-free.
7820A GC Installation Guide Connect the cabling Connect the power cord to the GC and the wall outlet. Figure Connect the LAN cable (part number 8121-0940) between the GC and PC. Installation requires a direct connection. Do not connect to a site LAN, hub, or switch for initial checkout.
7820A GC Installation Guide 7693A Automatic Liquid Sampler installation If available, install the 7693A injector as shown below. Install the G4513-20561 mounting posts on the inlet cover plate. See Figure Figure 9 Mounting post installation Install the injector on the mounting post. See Figure The GC may include a small cover plate, G3450-00152.
Page 19
7820A GC Installation Guide ALS PORT TRAY BACK FRONT Figure 10 Sampler installation and cable connection Connect the sampler control cable. See “Cables and Back Panel Connectors" on page 32 for more information about making connections to the GC. If using a G4513A injector, prepare it for use. (See the 7693A Installation, Operation, and Maintenance manual for instructions provided on the sampler.)
7820A GC Installation Guide Prepare for checkout Complete checkout instructions for all inlets and N OTE detectors are available in the 7820 Operation manual. Prepare the checkout sample. FID: 5188-5372 TCD: 18710-60170 NPD: 18789-60060 FPD: 5188-5953 uECD: 18713–60040 or 5183-0379 (Japan) If using a sampler, prepare solvent and waste vials.
7820A GC Installation Guide 4 mL Isooctane (FID, uECD, NPD, FPD) Hexane (TCD) Empty Turn on the GC. If purchased, install the NPD bead and configure the NPD. Refer to the 7820A Maintenance manual or the instructions that come with the bead.
Page 22
7820A GC Installation Guide The installation requires that you connect directly to the GC using this address. However, after installation you should change the GC IP address or set it to use DHCP. If needed, install the PC. Record the PC's current LAN address (either the IP address, subnet mask, and gateway;...
GC and GC/MSD Hardware User Information & Utilities DVD. Click on Install GC and GC/MSD Documentation, then follow the instructions for installing the online manuals. Installing the 7820A GC documentation also installs the software keypad. 7820A GC Installation Manual...
Page 24
7820A GC Installation Guide Install the Instrument Utility software. After installation completes, let the installer launch the Instrument Utility. 7820A GC Installation Manual...
7820A GC Installation Guide Run the Installation Wizard When run for the first time, the Instrument Utility will open to the Configure Instruments window. Select Instrument Type Agilent 7820. The default IP address (192.168.0.26) displays. Click Add Instrument to connect to the GC. Close the Configuration window.
Page 26
7820A GC Installation Guide Select your detector from the list. Select the checkout test for the detector, then click Run Test Now. • If using an FPD, refer instead to the Advanced User Guide. Perform the procedure as described in the manual for the checkout sample supplied with your instrument.
Page 27
7820A GC Installation Guide Select the checkout test for the detector. For example, Front FID Checkout Test. 7820A GC Installation Manual...
Page 28
7820A GC Installation Guide The test will prompt you to first purge and bake out the GC. Perform the bakeout. Watch the detector’s signal output. When bakeout ends, the detector output should meet the criteria listed below. Stable output < 20 pA Stable output between 12.5 and 750 µV...
Page 29
7820A GC Installation Guide Print or create a PDF version Send the results to Agilent. At this point, GC installation is complete. 7820A GC Installation Manual...
7820A GC Installation Guide Set Permanent GC and PC IP Addresses Your PC and GC are currently connected directly through a LAN cable. If this PC runs your data system, and you have no other GCs, you can leave the system as-is.
7820A GC Installation Guide Install the Data System The PC and GC are now installed and ready for use on your site LAN. The next step is to install your Agilent data system. Install the software, configure it to communicate with the GC, then make sure it is operating correctly.
7820A GC Installation Guide Cables and Back Panel Connectors Some parts of an analysis system are connected to the GC by cables. These cables and the back panel connectors to which they connect are described in this section. Figure 12...
7820A GC Installation Guide Sampler connectors If using an ALS, connect to the GC using the following connectors: ALS PORT Tray The sampler tray, if equipped. ALS PORT Front An injector, typically mounted over the front inlet. ALS PORT Back An injector, typically mounted over the back inlet.
7820A GC Installation Guide Using the Remote Start/Stop cable Remote start/stop is used to synchronize two or more instruments. For example, you might connect an integrator and the GC so that the [Start]/[Stop] buttons on either instrument control both of them. You can synchronize a maximum of ten instruments using Remote cables.
7820A GC Installation Guide The bus is passively pulled high. Leakage current out of a port must be less than 0.2 milliamps to keep the voltage from being pulled lower than 2.2 volts. Higher leakage current may cause the state to be interpreted as a low.
Start (Low True) Request to start run/timetable. Receiver is any module performing runtime-controlled activities. The 7820A GC requires a pulse duration of at least 500 micro-seconds to sense a start from an external device. 7820A GC Installation Manual...
7820A GC Installation Guide Start Relay (Contact Closure) A 120 millisecond contact closure – used as an isolated output to start another device that is not compatible or connected with APG Remote pin 3. Stop (Low True) Request to reach system ready state as soon as possible (for example, stop run, abort or finish, and stop injection).
7820A GC Installation Guide Connecting Cables Connect a GC to an Agilent data system computer using LAN communications using a LAN cable. See Figure 15 below. LAN cable 8121-0940 LAN hub/switch Crossover LAN cable 5183-4648 LAN cable LAN cable 8121-0940...
7820A GC Installation Guide Cable Diagrams Analog cable, general use Connector 1 Connector 2 The pin assignments for the general use analog out cable are listed in Table Table 3 Analog cable, general use, output connections Connector 1 Connector 2, wire color...
7820A GC Installation Guide Remote start/stop cable Connector 1 Connector 2 The pin assignments for the remote start/stop cable are listed in Table Table 4 Remote start/stop cable connections Connector 1, Connector 2, wire color Signal 9-pin male Black Digital ground...
7820A GC Installation Guide For More Information The GC and data system installation is now complete. For more information refer to: • GC Operating Guide for familiarization and everyday operating instructions • GC Safety guide 7820A GC Installation Manual...