Maintaining the GC/MSD Interface

The GC/MSD interface requires no periodic maintenance
Rarely, the heater cartridge in the GC/MSD interface fails. In those cases, it is
necessary to replace the heater and sensor. This section contains procedures
for removing the heater and sensor and installing new ones. See
Analyzer chamber
Ionization chamber
Heater/sensor assembly
Figure 27
EI GC/MSD interface
5975 Series MSD Troubleshooting and Maintenance Manual
General Maintenance
Figure
28.
Insulation
Heater sleeve
MSD
GC oven
4
153