Touch Panel (Touchscreen) Calibration
The 4VetJ touchscreen has a default calibration.
If the system malfunctions, it is possible that the touchscreen
has become uncalibrated. You can suspect that this is the
case if the accuracy is low when a measurement is made,
for example.
To calibrate the panel, enter the Settings menu and choose
the /Display/ tab. Press the [Calibration touch panel]
button.
The system shows the message: "Do you want to allow this
app to make changes to your device?." Follow these steps:
1. Press [Yes]. A window with the touch panel settings is
shown.
2. A Touch Panel Device List is shown. On the left of the
window, buttons for Device Addition, Basic Setting,
Touch Setting, Mouse Setting, and Tool Setting are
shown. List Update, and Monitor configuration options
are shown beneath the Device List. Select the bookmark
[Basic settings] on the left of the window.
3. In the section "Calibration" press the button [4Point].
4. The software shows four points in a sequence, which
should be touched in the middle.
5. When the last point is touched, an [OK] button appears.
Press it to end calibration.
6. Close the calibration software by pressing [Exit] in the
bottom left corner of the window.

Audio

The /Audio/ tab is used to set the level of the sound volume.
To test the level of sound volume, press [Test Doppler].

Measurements

The /Measures/ tab is used to modify the list of the displayed
parameters in the Aging tables.
To stop displaying these measurement, deselect the field.
Chapter 2 4VetJ Settings and Functions