Choosing A Location - Invacare HomeFill II User Manual

Oxygen system
Hide thumbs Also See for HomeFill II:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Invacare® HomeFill® II Oxygen System

3.3 Choosing a Location

DANGER!
Risk Of Death, Injury, Or Damage From Fire
Textiles, oil or petroleum substances, grease, greasy
substances and other combustibles are easily ignited
and burn with great intensity in oxygen enriched air and
when in contact with oxygen under pressure. To avoid
fire, death, injury or damage:
– DO NOT SMOKE while using this device.
– DO NOT use near OPEN FLAME or IGNITION
SOURCES.
– DO NOT use any lubricants on the compressor unless
recommended by Invacare.
– NO SMOKING signs should be prominently displayed.
– Avoid creation of any spark near oxygen equipment.
This includes sparks from static electricity created by
any type of friction.
– Keep all matches, lighted cigarettes or other sources
of ignition out of the room in which this compressor
is located and away from where oxygen is being
delivered.
– Keep the oxygen tubing, cord, and compressor out
from under such items as blankets, bed coverings,
chair cushions, clothing, and away from heated or hot
surfaces including space heaters, stoves, and similar
electrical appliances.
The compressor should be at least three inches (7.6 cm) away from
walls, draperies or furniture to ensure sufficient air flow.
Avoid placing the compressor next to heaters, radiators or hot air
registers.
DO NOT use in a closet.
14
The compressor should sit on a hard flat stable surface, such as a
table or the Ready Rack.
The HomeFill Ready Rack is sold separately and used to
position the HomeFill compressor on top of a Perfecto
Platinum oxygen concentrator. Refer to the HomeFill Ready
Rack instruction sheet part numbers 1148363 and 1134824
for more information.
or
2
1145804-F-00

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ioh200aw

Table of Contents