Siemens SIWAREX FTA Device Manual page 164

Weighing electronics for automatic scale
Table of Contents

Advertisement

Messages and Diagnostics
Error
Technology message - description
No.
12
Timeout print-out
13
Error Output device not ready
14
Logging not possible since the weight
lies outside of the weighing range
15
16
Logging not possible, Protokoll-ID error Logging is not possible because no protocol
17
Logging is not possible, printer not ready Logging is not possible because the printer
18
Logging is not possible, MMC is
defective
19
Logging is not possible, SIWATOOL
program is active
20
21-40 Weighing function error
21
Weighing command thrown out for lack
of a standstill
22
Zero setting is not possible
23
Zero setting / taring not possible
24
25
26
27
7-152
Solution
A print job was not complete within the
predefined monitor time for logging in DR4.
Check the output device or increase the
monitoring time.
Logging not possible because the defined
output device is not ready.
If the printer is defined as the output device,
check whether the printer is connected
properly and that it is installed and ready for
operation.
If the output device is selected as the MMC,
check for whether a formatted MMC is
inserted.
Logging is not possible because the weight lie
outside of the weighing range for a
calibratable application (Code in DR3 OIML).
Reserved
ID was defined in the protocol string
(calibratable application, Code in DR3 -
OIML).
has not indicated that it is ready. Check the
printer.
Logging is not possible because the MMC is
defective or has been removed. Change the
MMC or insert a formatted MMC.
Logging is not possible because the
SIWATOOL program is active on the
interface. Disconnect the connection to the
PC and connect a printer.
Reserved
Taring/zero setting or logging is not possible
because no standstill was achieved within
standstill time 1 in DR 3.
The start-up weight lies outside of the defined
value range defined in DR 3 for the maximum
positive and negative weight for zero setting.
Zero positions or zero setting not possible
because the gross lies outside the zero
setting range or taring not possible because
the tare maximum load has been exceeded.
Reserved
Reserved
Reserved
Reserved
SIWAREX FTA