6050H Calibration Bath
User's Guide
60
• If the instrument is used in a manner not in accordance with the equip-
ment design, the operation of the bath may be impaired or safety hazards
may arise.
NOTE: When checking the over-temperature cutout, be sure that the tem-
perature limits of the bath fluid are not exceeded. Exceeding the tempera-
ture limits of the bath fluid could cause harm to the operator, lab, and
instrument.
• The over-temperature cutout should be checked every 6 months to see that
it is working properly. In order to check the user selected cutout, follow
the controller directions (Section 9.8) for setting the cutout. Both the man-
ual and the auto reset option of the cutout should be checked. Set the bath
temperature higher than the cutout. Check to see if the display flashes cut-
out and the temperature is decreasing.