Service and Troubleshooting
November 2017
Host diagnostic message
Factory Approval Error
SIS Configuration Error
Function Not Supported
Antenna Type Configuration
Error
Factory Calibration Error
Analog Out Span Configuration
Error
Analog Out Calibration Error
SIS Multidrop Error
Engineering Unit Configuration
Error
84
Description
The Sensor Module factory approval is
missing.
The Transmitter factory approval is
missing.
It is currently not possible to enable Safety
Mode due to other active alerts.
Note
Rosemount 5408:SIS only supports liquids
level measurement when operating in
Safety (SIS) mode.
Functionality in the device is enabled, but
not supported by this device.
Additional features may be enabled by
purchasing an upgrade of the device.
The configured Antenna Type is not
supported by the device.
The factory calibration in the device is
missing.
The span for the configured analog out
range is too small.
Analog output calibration failed.
®
HART
multidrop mode is not supported
for safety (SIS) devices. Only 4-20 mA
output is supported for safety devices.
One of the configured engineering units is
not supported by the device.
00809-0100-4408, Rev BA
Recommended actions
1. Restart the device.
2. Restore default settings and reconfigure device.
3. If the condition persists, replace the device.
1. Clear other active alerts by priority order until this
alert is cleared.
2. Change Operational Mode to Control/Monitoring
if device is not intended to be used as safety
device.
3. If the condition persists, restore default settings
and reconfigure device.
1. Check that selections for variables (e.g. Primary
Variable) are supported by this device.
2. Turn off functionality not supported by this
device.
3. Consider purchasing an upgrade of the device to
access additional variables and functionality.
4. If condition persists, restore default settings and
reconfigure device.
1. Check configuration of Antenna Type.
2. Make sure the configured antenna type matches
the physical antenna for the device.
1. Replace the device.
1. Increase analog out span by adjusting Upper or
Lower Range Value.
1. Try calibrating the analog output again.
2. If the condition persists, replace the device.
1. Disable multidrop mode.
2. Change Operational Mode to Control/Monitoring
if device is not intended to be used as safety
device.
3. If the condition persists, restore default settings
and reconfigure device.
1. Check unit configuration.
2. If condition persists, restore default settings and
reconfigure device.
Reference Manual
Service and Troubleshooting