Emerson Rosemount 3051P Reference Manual page 70

In-line pressure, with hart protocol
Hide thumbs Also See for Rosemount 3051P:
Table of Contents

Advertisement

Troubleshooting
May 2017
Table 6-3. Status: Maintenance – Fix Soon
LCD
Alert name
display
screen
Sensor
Temperature
TEMP
Beyond Limits
LIMITS
Electronics
Temperature
Beyond Limits
MEMRY
Electronics
WARN
Board
(also in
Parameter
advisory)
Error
Configuration
STUCK
Buttons
BUTTON
Operator Error
Table 6-4. Status: Advisory
LCD
Alert name
display
screen
Non-Critical
User Data
Warning
MEMRY
WARN
Sensor
Parameter
Warning
LCD Display
[If display
Update
is not
Failure
updating]
[none]
Configurati
on Changed
60
LOI screen
The sensor temperature has
exceeded its safe operating
TEMP OUT
LIMITS
The temperature of the
electronics has exceeded its
safe operating range.
MEMORY
A device parameter does not
WARN
match the expected value.
(also in
The error does not affect
advisory)
transmitter operation or
analog output.
STUCK
Device is not responding to
BUTTON
button presses.
LOI
Problem
screen
A user written parameter
does not match expected
value.
MEMORY
WARN
A user written parameter
does not match expected
value.
[If display
The LCD display is not
is not
receiving updates from
updating]
the pressure sensor.
A recent change has been
[none]
made the device by a
secondary HART
such as a handheld device.
Problem
1. Check the process and ambient conditions are
range
2. Replace the pressure transmitter.
1. Confirm electronics temperature is within limits
2. Replace the pressure transmitter.
1. Replace the pressure transmitter.
1. Check configuration buttons are not stuck.
2. Replace the pressure transmitter.
1. Confirm and correct all parameters listed in Device
Information.
2. Perform a Device Reset.
3. Replace the pressure transmitter.
1. Confirm and correct all parameters listed in Device
Information.
2. Perform a Device Reset.
3. Replace pressure transmitter.
1. Check the connection between the LCD display and the
circuit board.
2. Replace the LCD display.
3. Replace the pressure transmitter.
1. Verify that the configuration change of the device was
intended and expected.
2. Clear this alert by selecting Clear Configuration
Changed Status.
®
master
3. Connect a HART master such as AMS Device Manager or
similar which will automatically clear it.
Reference Manual
00809-0700-4001, Rev AA
Recommended action
within –85 to 194 °F (–65 to 90 °C).
of –85 to 194 °F (–65 to 90 °C).
Recommended action
Troubleshooting