Accuracy Specifications - Fluke BT520 User Manual

Battery analyzer
Hide thumbs Also See for BT520:
Table of Contents

Advertisement

BT520
Users Manual

Accuracy Specifications

Accuracy is specified for a period of one year after calibration, at 18 °C to 28 °C (64 °F
to 82 °F), with relative humidity to 80 %. Accuracy specifications are given as: ±([% of
reading] + [number of least significant digits]). Accuracy specification assumes ambient
temperature stable ±1 °C.
Battery Internal Resistance
V ac (45 Hz to 500 Hz with low-pass
filter)
Frequency (Display with V ac)
Trigger level: ≥ 10 mV @V ac;
AC Voltage Ripple (20 kHz max)
[1]
The measurement is based on ac injection method. The injected source signal is <100 mA, 1 kHz.
1-12
Function
3 mΩ
30 mΩ
[1]
300 mΩ
3000 mΩ
6 V
V dc
60 V
600 V
600 V
45 Hz to 500 Hz
600 mV
6000 mV
Range
Resolution
0.001 mΩ
0.01 mΩ
0.1 mΩ
1 mΩ
0.001 V
0.01 V
0.1 V
0.1 V
0.1 Hz
0.1 mV
1 mV
Accuracy
1 % + 8
0.8 % + 6
0.8 % + 6
0.8 % + 6
0.09 % + 5
2 % + 10
0.5 % + 8
3 % + 20
3 % + 10