Detector Handling

The device contains sensitive electronics that are susceptible to vibration, shock, drop, and impact.
When handling the device, use the handle and/or use both hands to manipulate the detector into the
correct anatomical position for the exam.
WARNING
Do not swing the device into hard surfaces, especially corners, as this may scratch the
cover, create an image quality artifacts, or damage the electronics inside.
Special Instructions if the detector is dropped: Inspect the exterior for any possible cracks. Run
Detector Check to test the detector. It will be clear if the electronics are not functioning. Other possible
failures may include communication problems, image quality degradation, and loss of power. If any or all
of these occur, call your GE Service Representative.
x
Do not drop objects onto the detector.
x
Do not use the detector as a stretcher to lift a patient.
x
Do not drop the detector at any time.
x
Do not prop the device on an edge, against wall or bed. Keep detector in cradle, bucky, or other GE-
supplied container.
x
Do not place other objects or patients on the detector if it is not on a flat surface, as shown in .
x
Do not use unapproved chemical cleaners.
x
Do not immerse detector into water or other liquids.
x
Do not use a defibrillator while patient remains in contact with detector.
x
Do not place objects on detector if it is not lying flat
Detector surfaces have been treated with a finish to provide a smooth and easily cleanable surface. Take
care to protect the surface from scratches.
WARNING
Extra precautions should be taken if the device will be exposed to excessive amounts
of bodily fluids or liquids.
The detector is designed to prevent some liquids or particulate matter from getting inside the cover. It
can sustain a temporary splash or spray, but it is not designed to be immersed in liquid (not even tempo-
rarily).
Optima XR220amx X-Ray System
Operator Manual
CTRL+SHIFT+F to search using Search Index (faster)
54400222-1EN
Rev 1 DRAFT 11092010
© 2010 General Electric Company. All rights reserved.
REVIEW COPY
Chapter 13: Digital Detector
13-2

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents