1.4.4 Minimum elevator depth requirements
See
1.4.1 Product dimensions and weights
Section 1.5 Product storage and handling requirements
Refer to
atmospheric pressure requirements.
The maximum incline that the system can navigate is 7 degrees, and the maximum threshold height
over which the system can drive is 2.5 cm (1 in).
The system is not designed to travel over rough or bumpy ground. Take care when moving
the system to avoid sharp bumps, or steps larger than the 2.5 cm (1 in) threshold limit.
Section 1.6 Pre-installation results
After the pre-installation work is completed, you should be able to check off each of these items:
1.
2.
3.
4.
5.
6.
7.
8.
9.
10. If tablet holder option is purchased, the tablet is ready to install.
Page 24
3
EVISION
Chapter 3 Environmental (HVAC) requirements
Radiation physicist consulted about radiation protection
Electrical service in place at the ratings specified
Charging/downloading area size adequate for equipment
Secure location to store loose items identified
Network requirements met or exceeded
If wireless option is purchased, wireless network capabilities met
Network/workflow information collected
IP addresses for DICOM and broadband identified
Broadband access provided for InSite ExC connection
Section 1.5 Product storage and handling requirements
O
and
1.4.2 Shipping dimensions and
for temperature, humidity, and
XR200
/XR220
PTIMA
AMX
AMX
P
-I
M
RE
NSTALLATION
ANUAL
weights.