Installation; Basic System Requirements - Hach BioTector B7000 User Manual

Online toc tn analyzer
Table of Contents

Advertisement

Section 5
5.1

Basic System Requirements

Power and Signal Requirements
Mains Connection:
Mains Wire Specification:
Signal Wire Specification:
Power Consumption:
Electrical Connections:
Carrier Gas and Reagent Requirements
Carrier Gas (Oxygen) Requirements
Oxygen Quality:
Oxygen Supply Pressure:
Oxygen Usage:
Reagent Requirements
1.8 N Sulfuric Acid (H
1.2 N Sodium Hydroxide (NaOH),
0.04 N Hydrochloric Acid Solution (HCl Water for FMI Heavy Duty Circulation Pump systems only),
TN Cleaning Solution (a mixture of 0.5 N Hydrochloric Acid and 0.042 M Sodium Oxalate),
Deionised Water (DI Water).
Sample, Drain and Exhaust Requirements
Sample Inlet & Outlet Pressure:
Sample Inlet Temperature:
Sample Flow Rate:
Sample Particle Size:
Drain & Exhaust:

Installation

115V AC, 60Hz or 230V AC, 50Hz (10%)
Number of Cores = 3
Current Rating minimum = 10 Amps
CSA (Cross Sectional Area minimum) = 1.50mm
Number of Cores = 10 (+2 cores per additional signal)
Current Rating minimum = 1 Amp
CSA (Cross Sectional Area minimum) = 0.22mm
Maximum 300 W (VA)
Typically 5 cable glands, PG13.5, clamping range 6 - 12 mm
Oxygen, free of CO
Typical oxygen purity in the cylinder is greater than >99.5%
Typical oxygen purity from the concentrator is 93% (±3%) with
balance gas Argon
Option A: 1250-2000 mbar from the oxygen concentrator with
Option B: Depending on the type, from 550 mbar to 750 mbar from
Option C: 1000 mbar from the welding grade oxygen cylinder
22 L/hour (367 ml/min) average
SO
),
2
4
, CO, Nitrogen, Hydrocarbons and Water
2
instrument air
the oxygen concentrator with compressor
Ambient
2°C – 60°C (36°F - 140°F)
Minimum 100ml per sample
Up to 2 mm Ø, soft particulates
Ambient
2
2
Page 49