4. PREPARATION AND SAFETY CHECK

4-1 PREPARATION

WARNING - Pentax video processors are electro-medical devices incorporating delicate components and sophisticated
circuitry that should not be subjected to harsh conditions, including excessive vibrations and/or severe impact. Never drop this
equipment or subject it to severe impact as it could compromise the functionality and /or safety of the unit. Should this
equipment be mishandled or dropped, do not use it. Return it to an authorized Pentax service facility for inspection and repair.

(1) SETTING UP THE EPK-700

1. Place the EPK-700 video processor on a stable, level surface (cart, counter, stand,etc..).
NOTE
• Avoid places where the EPK-700 may be splashed with liquid.
• Absolutely DO NOT use in any environment with explosive or flammable gases.
• Do not block the venting grids on the sides of the video processor.
• Do not install, operate or srore electro-medical equipment in a dusty environment.
Accumulation of dust within these units may cause malfunction, smoke, or ignition.
• When moving the EPK-700, do not hold the scope locking lever.
2. Make sure the power switch is OFF.
3. Plug the power cord into the power source
using the three (3) prong plug supplied with the unit.
4. Ensure the keyboard is connected properly.
CAUTION
- To prevent electric shock, connect power cords of
ancillary equipment to medical isolation transformers.
NOTE
-When using medical isolation transformers, be sure to
check that the total power consumption of all the devices does
not exceed the isolation transformers' power ratings. Make
sure that the power cord is connected to the main with a three
(3)prong plug (In the U.S.A. use UL2601-1 rated isolation
transformers and/or power strips only).

(2) CONNECTING THE WATER BOTTLE

1. Fill the water bottle approximately 2/3 full with sterile water.
2. Screw the water bottle cap assembly to the water bottle
snugly.
NOTE
-Do not over tighten the water bottle cap
3. Set the Air/Water- Drain lever to A/W .
4. Insert the water bottle air pipe stem into the EPK-700 water
bottle receptacle and press until the water bottle 'clicks' into
position.
NOTE
-Do not press the water bottle too forcefully into the
EPK-700. Rough handling may cause water to leak onto/into
the video processor.
5. Insert the Air/Water connector into the holder on the water
bottle cap assembly until the endoscope is connected
NOTE
-Always disconnect the water bottle before moving the
video processor into a position not common to normal use.
Always disconnect the water bottle before packing the EPK-
700 for shipment.
14
RGB OUTPUT
Y / C-OUTPUT
Y / C-OUTPUT
RGB OUTPUT
BREAKER (S) 2A
VIDEO
VIDEO
VIDEO PROCESSOR
MODEL EPKC-700P
ASAHI OPTICAL CO., LTD.
INPUT 100V~ 50-60Hz 1.8A
MADE IN JAPAN
RS-232-C
CONTROL1
CONTROL2
SERIAL NO.
KEY BOARD
WARNING
LIFE METER
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,
REPLACE ONLY WITH SAME TYPE AND RATING OF FUSE.
A / W
DRAIN