Specifications - Hach APA 6000 Operation Manual

High and low range copper process analyzer
Table of Contents

Advertisement

Specifications

Performance
Sample Requirements
Page 4
Specifications
Specifications are subject to change without notice.
See instrument-specific specifications in the APA 6000™ Installation and
Maintenance Manual, Cat. No. 62005-18.
Range: 0.05 to 10 mg/L Cu
Accuracy:
Low range (optimum from 0.03 to 1.00 mg/L): ±3% or ±0.03 mg/L,
whichever is greater.
High range (optimum from 1 to 10 mg/L): ±5% or ±0.05 mg/L,
whichever is greater.
Repeatability: ±3% of reading or ±0.03 mg/L, whichever is greater
Response Time: For single channel instruments, less than 5 minutes for 90%
response to step change at sample inlet.
Cycle Time: 4.5 minutes (average)
Sample Temperature Range: 5 to 40 °C (41 to 104 °F)
Instrument Detection Limit: 0.30 mg/L or less
Operating Temperature: 5 to 40 °C (41 to 104 °F)
Relative Humidity: 5 to 95% relative humidity, non-condensing
Sample Pressure: 0.5 to 30.0 psig (0.03 to 2.04 bar)
Sample Temperature: 5 to 50 °C (41 to 104 °F)
Sample Flow: 100 to 2000 mL/min. maximum
Sample Inputs: Up to two sample streams