Safety Precautions Please read this entire manual before unpacking, setting up, or operating this instrument. Pay particular attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage the equipment. To ensure the protection provided by this equipment is not impaired, do not use or install this equipment in any manner other than that which is specified in this manual.
Specifications Specifications are subject to change without notice. See instrument-specific specifications in the APA 6000™ Installation and Maintenance Manual, Cat. No. 62005-18. Performance Range: 0.05 to 10 mg/L Cu Accuracy: Low range (optimum from 0.03 to 1.00 mg/L): ±3% or ±0.03 mg/L, whichever is greater.
Operation DANGER Handling chemical samples, standards, and reagents can be dangerous. Review the necessary Material Safety Data Sheets and become familiar with all safety procedures before handling any chemicals. DANGER La manipulation des échantillons chimiques, étalons et réactifs peut être dangereuse. Lire les Fiches de Données de Sécurité...
Instrument Overview 1.1 How to Use the Manuals This instrument requires two manuals for complete operating instructions. The APA 6000™ Installation and Maintenance Manual (Cat. No. 62005-18) provides instructions for instrument installation, menu setup, operation, maintenance, and troubleshooting. This Parameter-specific Operation Manual (Cat. No. 62029-18) contains information related to parameter specific instrument components and software functions.
Tefzel. The small diameter allows for low-volume, high-flow rates. The higher flow rate, along with properly applied sample conditioning, prevents the tubing from plugging. General information for replacing the tubing is presented in the APA 6000 Installation and Maintenance Manual (Cat. No. 62005-18). CAUTION...
Section 1 VORISCHT Es wird dringend empfohlen, die Sicherheitsdatenblätter vor der Handhabung von Behältern, Tanks und Zufuhrsystemen, die chemische Reagenzien und Standardsubstanzen enthalten, aufmerksam durchzulesen, damit Sie sich mit den beim Umgang mit diesen Chemikalien notwendigen Vorsichtsmaßnahmen, Risiken und Notfallschutzmaßnahmen vertraut machen, Es wird empfohlen, in allen Situationen, in denen mit einem Kontakt mit Chemikalien zu rechnen ist, eine Schutzbrille zu tragen.
Page 9
Section 1 Figure 1 Plumbing Diagram for a Copper APA Analyzer Table 1 Replacement Tubing Lengths for Figure 1 Item Description Length From... To... Tefzel tubing, 0.03 ID x 0.062 OD 7.3 in. (18.5 cm) Port 1 Colorimeter in Tefzel tubing, 0.03 ID x0.062 OD 24 in.
Section 2 Theory of Operation The APA 6000™ analyzer is designed to provide reproducible results through consistent solution handling. Every step in an analysis is performed the same way each time, ensuring accurate results. Accuracy can be verified and established by measuring grab samples with a known copper concentration.
Section 2 2.2 Reagent and Standard Consumption The rate of reagent use depends on many factors, including the number of times the instrument cycles, calibration frequency, and the number of clean cycles. Table 3 provides the approximate volume (in mL) of the reagents consumed with each type of function per cycle.
Section 3 Parameter Specific Menu Functions 3.1 Calibration History This option allows you to review past calibration data. 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4.
Section 5 Bench Comparison Testing 5.1 Comparable Bench Methods The procedure presented here uses the APA 6000™ Copper Reagents to create a correlation between the analyzer and a method performed in the laboratory. Other copper procedures may be used to create that correlation. Other methods presented in Standard Methods for the Examination of Water and Wastewater should provide comparable results.
Page 18
Repeat Steps 2–4, Using a 1-cm sample Within 10 minutes of Record the value. using the APA Standard 2 cell and a Hach DR/4000* adding reagents, transfer (low range) 1.0 mg/L Spectrophotometer, 3 mL of the 0 ppm in Step 2 and 3 in place of zero at 440 nm with standard into a 1-cm cell.
Section 5 Testing Samples 440 nm Using a volumetric Add 5 mL of APA Within 10 minutes of Calculate the total pipet, transfer 5 mL of Copper Reagent 1 (buffer) adding reagents, transfer copper concentration in APA Copper Reagent 2 to the same flask.
APA 6000™ Copper Standard 1, 0.0 mg/L .................. 1 L..27562-53 APA 6000™Copper Standard 2, low range, 1.0 mg/L ..............1 L..27563-53 APA 6000™ Copper Standard 2, high range, 10.0 mg/L ............. 1 L..27564-53 APA 6000™ Cleaning Solution..................... 1 L..26964-53 Required Reagents for the Bench Procedure APA 6000™...