Section 1
Performance specifications
Operating mode
Colorimetric evaluation
Source lamp
Wavelength range
Wavelength accuracy
Wavelength reproducibility
Wavelength resolution
Wavelength calibration
Wavelength range for color
measurement
Scanning speed
Spectral bandwidth
Photometric measuring range
Photometric accuracy
Photometric linearity
Stray light
Data log
Physical and environmental specifications
Width
Height
Depth
Weight
Operating requirements
Storage requirements
Additional technical data
Mains connection
Interfaces
Enclosure rating
Protection Class
Specifications
Specifications are subject to change without notice.
Color measurement, color difference measurement, Absorbance and
Concentration
All color values calculated for standard illuminant C of ASTM E 308 and a
2°standard observer.
Gas-filled Tungsten (visible)
320–1100 nm
± 1.5 nm (wavelength range 340–900 nm)
≤ 0.1 nm
1 nm
Automatic
380 to 720 nm step 10 nm
≥ 12 nm/s (in steps of 1nm)
5 nm
± 3.5 Abs (wavelength range 340–900 nm)
5 m Abs at 0.0 to 0.5 Abs
1% at 0.50 to 2.0 Abs
< 0.5% to 2 Abs
< = 1% at > 2 Abs with neutral glass at 546 nm
< 0.1% T at 340 nm with NaNO
500 color measurements, 50 color reference values, 500 photometric
measurements, 20 wavelength scans, 20 time scans
368 mm (14.5 in.)
144 mm (5.7 in.)
359 mm (14.1 in.)
6.4 kg (14.11 Ib)
10–40 °C (50–104 °F), max. 80% relative humidity (non-condensing)
–40–60 °C (-40–140 °F) max. 80% relative humidity (non-condensing)
External power supply: 100-240V/50-60Hz (Input); 15V/30VA (output)
Use only screened cables with maximum length of 3 meters.
1 x USB type A
1 x USB type B
IP3X
Class II
2
5