Contrast Calibration Adjustment - Fluke 435-II Service Manual

Three phase energy and power quality analyzer
Hide thumbs Also See for 435-II:
Table of Contents

Advertisement

Fluke 434-II/435-II/437-II
Service Manual
Functions of the keys F1-F4 are:
F1
F2
F3
F5

5.4 Contrast Calibration Adjustment

After entering the calibration mode the display shows:
WarmingUp (CL 0200):IDLE (invalid)
Do not press
calibration mode again.
Proceed as follows to adjust the maximum display darkness (CL 0100), the default
contrast (CL 0110) , and the maximum display brightness (CL 0120).
1.
2. Press
3. Using
4.
5.
6. Press
7. Using
8.
9.
10. Press
11. Using
12. Press
13. Now do the complete calibration. Press
5-6
select the previous step
PREV
select the next step
NEXT
start the calibration adjustment of the actual step
CAL
leave the calibration mode
EXIT
F3
now! If you did, turn the Analyzer off and on, and enter the
F1
Press
three times to select maximum darkness calibration
F3
CALIBRATE
adjust the display to the maximum darkness at which the test
pattern is only just visible (up = lighter, down = darker).
F2
Press
to return to the soft key menu.
F2
Press
to select default contrast calibration
F3
CALIBRATE
set the display to optimal (becomes default) contrast (up = lighter,
down = darker).
F2
Press
to return to the soft key menu.
F2
Press
to select maximum brightness calibration
F3
CALIBRATE
adjust the display to the maximum brightness, at which the test
pattern is only just visible.
F2
to return to the soft key menu (up = lighter, down = darker).
and continue at Section 5.5.
. The display will show a dark test pattern, see Figure 5-2
. The display shows the test pattern at default contrast.
. The display shows a bright test pattern.
F2
to select the next step (Warming Up),
Contrast (CL 0100):
Contrast (CL 0110):
Contrast (CL 0120):