5.5 Warming Up

The Warming Up state will be entered after entering the calibration mode (section 5.3),
or after selecting the next step if you have done the Contrast Calibration step CL 120
(section 5.4). The display will show
WarmingUp (CL 0200):IDLE (valid)
Unless you want to calibrate the display contrast only, you must always start the
calibration adjustment at the
make the calibration invalid!
The Warming Up consists of a 30 minutes warming-up period.
Proceed as follows to do the Warming Up:
1. Remove all input connections from the Analyzer.
F3
2. Press
The display shows the calibration step in progress, and its status.
The display shows
period is counted down from 00:29:59 to 00:00:00
3. Wait until the display shows
If you turn off the Analyzer now by accident, turn it on again immediately; now you
can select the calibration mode, and continue with step 4 below (press
several times, see 5.6).
If you turn off the instrument now, and you do not turn on immediately, the Analyzer
has cooled down, and you must repeat the Warming Up (select the calibration mode
and start at CL 0200).
F2
4. Press
Figure 5-2. Version & Calibration data.
or
Warming Up (CL 0200)
to start the Warming-Up.
WarmingUp (CL 0200)
Warmingup: READY
and continue at Section 5.6.
NEXT
(invalid).
step. Starting at another step will
:BUSY 00:29:59
Calibration Adjustment

5.5 Warming Up

The warming-up
.
F2
NEXT
5
5-7