Emerson Oxymitter 5000 Instruction Manual page 84

Oxygen transmitter with foundation fieldbus communications
Hide thumbs Also See for Oxymitter 5000:
Table of Contents

Advertisement

Instruction Manual
IB-106-350 Rev. 1.2
April 2001
6. Being careful not to disconnect the
board wiring, carefully lift the board and
spacer assembly from manifold (5) and
set aside. Do not lose o-rings (8) from
the bottom of spacers (9).
7. Tag and remove the leads from pres-
sure switch (12).
8. Place a 1-1/16 in. 6-point socket over
pressure switch (12) and remove.
When installing the pressure switch,
do not overtighten. Damage to the so-
lenoid may occur.
9. Install new pressure switch (12). Be
careful not to overtighten. Connect the
leads to the proper terminals on the
pressure switch.
10. Carefully install the board and spacer
assembly into manifold (5) by aligning
spacers (9) with the mounting holes on
the manifold and securing with screws
(11). Ensure o-rings (8) are installed
between the spacers and the manifold
surface.
11. Install manifold cover (14), and secure
with manifold cover lock (6) and screw
(7).
12. Turn on the calibration gases at the
cylinders.
e. Check Valve Replacement
Check valve (19, Figure 4-13) may stick or
become plugged over time. Replace when
necessary. If condensation deposits are
noted upon removal, consider insulating the
check valve.
f.
Pressure Regulator (Optional) Mainte-
nance
1. Pressure Adjustments. Reference air
pressure regulator (8, Figure 4-13) is
4-22
Maintenance and Service
factory set to 20 psi (138 kPa). Adjust
using the knob on top of the pressure
regulator if necessary.
Do not use fingers to release valve
stem. The valve may release air at
high pressures and cause injury.
2. Condensation Drain. To drain excess
moisture from the filter bowl of refer-
ence air pressure regulator (8), use a
screwdriver or comparable tool to peri-
odically release valve stem on the
bottom of the regulator.
g. Flowmeter Adjustments.
1. Calibration Gas Flowmeter. Calibration
gas flowmeter (17, Figure 4-13) regu-
lates the calibration gas flow and must
be set to 5 scfh. However, only adjust
the flowmeter to 5 scfh after placing a
new diffusion element on the end of the
Oxymitter 5000. Adjusting the flowme-
ter at any other time can pressurize the
cell and bias the calibration.
2. In applications with a heavy dust load-
ing, the O
may become plugged over time, caus-
ing a slower speed of response. The
best way to detect a plugged diffusion
element is to note the time it takes the
Oxymitter 5000 to return to the normal
process reading after the last calibra-
tion gas is removed and the calibration
gas line is blocked off. A plugged ele-
ment also can be indicated by a slightly
lower reading on the flowmeter.
Change the diffusion element when the
calibration gas flowmeter reads slightly
lower during calibration or when the re-
sponse time to the process flue gases
becomes very slow. Each time the dif-
fusion element is changed, reset the
calibration gas flowmeter to 5 scfh and
calibrate the Oxymitter 5000. For more
information on changing the diffusion
element, refer to paragraph 4-8.
Rosemount Analytical Inc.
A Division of Emerson Process Management
Oxymitter 5000
probe diffusion element
2

Advertisement

Table of Contents
loading

Table of Contents