Chapter 9. Useful Function

Temperature recording pen (Red)
Chart paper
Humidity recording pen (Blue)
Distance between time axes
ON
1
2
3
- 25 -

9-1. Pen trace o set compensation (POC)

Function overview

When recording the temperature and humidity on the chart
paper, those two traces inevitably have the di erent time
axes due to the positional di erence of the recording pens.
Use this function to compensate the distance and align the
time axes.
Reference
The table below shows the reference between the
chart speed and time required for the two traces to
synchronize (agree with each other) in time axis.
C
h
r a
s t
p
e
e
d
20mm/h
10mm/h
5mm/h
2.5mm/h

Setting POC

① Turn OFF ( 【
) the power switch.
② Turn ON the DIP switch No.2, then turn the power switch
ON again.
③ Temperature pen will soon start recording, while humidity
pen will wait for the time in the above Reference. Then it
will start recording.
CAUTION
Be sure to keep the DIP switch No. 5 to No. 8 OFF
state.

9-2. Marking function

If desired, press the 【MARK 】 key to leave a mark on the
temperature trace while recording.
CAUTION
No mark will be made if the recording pen goes out of
the chart paper.
9-3. Indicating dew point
temperature and absolute humidity
Press the 【ENTER /
DEW POINT/ABSOL. HUMIDITY
key during the normal measurement display to display the
dew point temperature and the absolute humidity for about
5 sec. At this time, 『D.P. 』 and 『g/m
i T
m
e
7.5 minutes approx.
15 minutes approx.
30 minutes approx.
1 hour approx.
』 will blink.