Hach HQ440d User Manual page 31

Hide thumbs Also See for HQ440d:
Table of Contents

Advertisement

Figure 14 Advanced report for sample data—2 lines
1 Probe model
2 Error message (if applicable)
3 Probe serial number (a "<" or ">" on the HQ40d
meter indicates the probe position)
Figure 15 Total report for sample data—3 lines
1 Method name for probe settings
2 Time of calibration, prefaced by "CAL" and
displayed as hh:mm in 24 h (or user-defined) format
3 Date of calibration (DD-MM-YY or user-defined
format)
Check standard reports
Check standard data is printed with 1 line of information
Figure 16 Check standard report
1 Report type (ST = check standard)
2 Report type (ST = check standard)
3 Time (hh:mm:ss in 24 h or user-defined format)
4 Date (DD-MM-YY or user-defined format)
Calibration reports
Calibration data is printed when the data log is sent to the printer or when probe data is sent to the
printer. Calibration data is printed with 2 lines of information
4 Additional units: shows all additional units
associated with the reading.
5 Probe settings: shows the highest-priority setting
associated with the reading
4 Calibration slope/ratio/constant
5 Offset—contents vary depending on type of
parameter and user settings. May be blank.
6 Operator ID: user-defined; shows "- - -" if undefined
(Figure
16)
5 Measured value
6 Units
7 Check standard status: Pass/Fail based on the
acceptance criteria
8 Operator ID: user-defined; shows "- - -" if undefined
(Figure
17).
English 29