The calibration parameters are accessed by pressing "SET" after the name of the pa-
rameter is displayed. The value of the parameter may be changed using the up or down
arrow. After the desired value is reached, press "SET" to set the parameter to the new
value. Pressing "EXIT" causes the parameter to be skipped ignoring any changes that
may have been made.
6.12.1
R0
This probe parameter refers to the resistance of the control probe at 0°C. The value of
this parameter is set at the factory for best instrument accuracy. The value ranges from
95 to 105. For values greater than 100.000, the display does not show the hundreds
placement. For values less than 100.000, the display shows the entire value. The R0
parameter is indicated by,
r0
00.014
To change the R0 setting, press the up or down arrows.
99.999
To accept the new setting, press "SET". Press "EXIT" to continue without storing the
new value.
S
6.12.2

ALPHA

This probe parameter refers to the average sensitivity of the probe between 0 and
100°C. The value of this parameter is set at the factory for best instrument accuracy.
aLpha
38530
To change the ALPHA setting, press the up or down arrows.
38600
To accept the new setting, press "SET". Press "EXIT" to continue without storing the
new value.
S
Flashes for one second and then the R0 setting is displayed
Current R0 setting (100.014)
New R0 setting
Accept the new R0 setting
Flashes for one second and then the ALPHA setting is
displayed
Current ALPHA setting
New ALPHA setting
Accept the new ALPHA setting