Fluke 9102S User Manual page 41

Dry-well calibrator
Hide thumbs Also See for 9102S:
Table of Contents

Advertisement

Command Description
Serial Interface Menu
Read serial sample setting
Set serial sampling setting to n seconds
Set serial duplex mode:
Set serial duplex mode to full
Set serial duplex mode to half
Set serial linefeed mode:
Set serial linefeed mode to on
Set serial linefeed mode to off
Calibration Menu
Read R0 calibration parameter
Set R0 calibration parameter to n
Read ALPHA calibration parameter
Set ALPHA calibration parameter to n
Read DELTA calibration parameter
Set DELTA calibration parameter
Miscellaneous Other Commands
Read firmware version number
Read structure of all commands
Read ALL operating parameters
Read set-point
Command
Format
sa[mple]
sa[mple]=n
du[plex]=f[ull]/
h[alf]
du[plex]=f[ull]
du[plex]=h[alf]
lf[eed]=on/of[f]
lf[eed]=on
lf[eed]=of[f]
r[0]
r[0]=n
al[pha]
al[pha]=n
de[lta]
de[lta]=n
*ver[sion]
h[elp]
all
*sr
Legend: [] Optional command data
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
Note: When DUPLEX is set to FULL and a command is sent to READ, the command is returned
followed by a carriage return and linefeed. Then the value is returned as indicated in the
RETURNED column.
Digital Communication Interface
Command
Example
Returned
sa
sa: 99999
sa=0
du=f
du=h
lf=on
lf=of
r
r0: 999.999
r=100.7
al
al: 9.99999999
al=0.003865
de
de=1.37
de: 9.99999
*ver
ver.9999x,9.99
h
list of
commands
all
list of
parameters
*sr
999.999 ohms
Interface Commands
Returned
Acceptable
Example
Values
sa: 1
0 to 10,000
FULL or HALF
ON or OFF
r0: 100.7
95.0 to 105.0
al: 0.003865
0.002 to 0.006
de: 1.50
0.0–3.0
ver.9102S,1.10
100.123 ohms
33