Agilent Technologies 54621A User Manual page 246

Hide thumbs Also See for 54621A:
Table of Contents

Advertisement

Performance Characteristics
Agilent 54620-series Performance Characteristics
* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and
±10 ° C from firmware calibration temperature.
Analog Channel Triggering
Range (Internal)
Sensitivity*
Coupling
Digital (D15 - D0) Channel Triggering (54621D and 5462 2D)
Threshold Range (used defined) ±8.0 V in 10 mV increments
Threshold Accuracy*
Predefined Thresholds
External (EXT) Triggering
Input Resistance
Input Impedance
Maximum Input
Range
Sensitivity
Coupling
Probe ID (Agilent/HP & Tek
Compatible)
7-8
±6 div from center screen
Greater of 0.35 div or 2.5 mV
AC (~3.5 Hz), DC, noise reject, HF reject and LF reject (~ 50 kHz)
±(100 mV + 3% of threshold setting)
TTL = 1.4 V, CMOS = 2.5 V, ECL = -1.3 V
1 M
±3%
~ 14 pF
CAT I 300 Vrms, 400 Vpk
CAT II 100 Vrms, 400 Vpk
with 10074C 10:1 probe:CAT I 500 Vpk, CAT II 400 Vpk
±10 V
dc to 25 MHz, < 75 mV
25 MHz to max bandwidth, < 150 mV
AC (~ 3.5 Hz), DC, noise reject, HF reject and LF reject (~ 50 kHz)