To capture glitches or narrow pulses with peak detect and infinite persistence
3 Press the
Display
key then press the
softkey.
Infinite persistence updates the display with new acquisitions but does not erase
previous acquisitions. New sample points are shown at normal intensity while
previous acquisitions are displayed in minimum intensity. Waveform persistence
is not kept beyond the display area boundary. Any time you change a waveform
setting such as sweep speed, delay time, vertical sensitivity, or offset, the
previous acquisitions will be erased, then begin to accumulate again.
Use infinite persistence to measure noise and jitter, to see the worst-case
extremes of varying waveforms, to look for timing violations, or to find
infrequent events.
Press the Clear Display softkey to erase previously acquired points. The display
will again accumulate points until Clear Display is pressed or
off.
4 Characterize the glitch with delayed sweep.
To characterize the glitch with delayed sweep, follow these steps:
• Press the Main/Delayed key, then press the Delayed softkey.
• To obtain a better resolution of the glitch, expand the time base.
Use the horizontal delay time knob (
expanded portion of the main sweep around the glitch.
To characterize the glitch, use the cursors or the automatic measurement
capabilities of the oscilloscope.
Making Measurements
Persist
(infinite persistence)
)to pan through the waveform to set the
Persist is turned
5-11