POST REPROCESSING AND STORAGE

WARNING
1)
Following reprocessing, the endoscope may either be reused or placed into storage.
2)
Prior to storage, ensure that all internal channels, endoscope components, instrument
surfaces, and accessories are thoroughly dry.
3)
The endoscope should be hung vertically in a clean, dry, well-ventilated storage
cabinet at room temperature. The insertion tube and light guide cable should be hung
and kept as straight as possible during storage.
4)
Prior to reuse, ensure that instrument has been properly inspected and fully prepared
for the next clinical procedure.
• Make sure that all removable components such as the suction control
valve, rubber inlet seal, etc., are detached from the endoscope. This will
allow for better air circulation through the internal channels and permit
thorough drying. All channels should be completely dry before storage.
• Never store the endoscope, its components, or accessories in the
carrying case, as this type of dark, humid, and unventilated environment
is conducive to bacterial colonization and increases the risk of cross
contamination. These cases are intended only for transportation of the
instrument, not storage.
• Never store the endoscope in areas of high humidity, high temperature,
or in direct exposure to sunlight or X-rays.
• Avoid storage of the endoscope in cabinets that have sharp edges,
exposed nails/screws, etc. Contact with sharp objects can puncture,
scratch, or otherwise damage the endoscope.
• When utilizing heated disinfectants for reprocessing PENTAX endoscopes,
the instruments should be allowed to return to room temperature prior
to use and/or further handling.