1
Endoscope
Sterilization
1-2-5-1-2. Preparation
WARNING
Consult the manufacturer of the sterilization tray regarding its directions for
use.
Prior to the sterilization using STERRAD
cap (OF-C5) is attached to the scope. Failure to do so can result in bursting of
the bending rubber.
1)
Detach cleaning adapter (OF-B155) and the inlet seal (OF-B190) are detached from
the scope.
(1)
Figure 1.28
2)
Detach PVE soaking cap (OE-C9) from the scope.
3)
Attach ventilation cap (OF-C5) to the scope
(1)
Figure 1.29
4)
Lay the APTIMAX Instrument Tray Mat on APTIMAX Instrument Tray
5)
Place the scope onto the mat as shown below
Figure 1.30
6)
Cover the tray with the lid.
®
NX™ system, ensure that Ventilation
(2)
(2)
– 37 –
(1) OF-B155
(2) OF-B190
(1) OF-C5
(2) OE-C9