Pentax EB-1970TK Instructions For Use Manual page 54

Video bronchoscopes
Table of Contents

Advertisement

1
Accessory
Sterilization
1-3-3-2-2. Preparation
WARNING
Contact the manufacturer of the tray regarding directions for use of the tray.
1)
Lay APTIMAX Instrument Tray Mat on APTIMAX Instrument Tray.
2)
Place the suction control valve (OF-B179) and the inlet seal (OF-B190).
3)
Cover the tray with the lid.
1-3-3-2-3. Wrapping
WARNING
• Contact the manufacturer regarding directions for use of the wrap and
tape.
• The suction control valve (OF-B179) and the inlet seal (OF-B190) may
be wrapped and sterilized together with the endoscope. In this case, do
not attach any components other than the ventilation cap (OF-C5) to the
endoscope.
• The ventilation cap (OF-C5) must be attached to the endoscope when
performing sterilization using STERRAD
• Secure the wrapping with a length of tape that is sufficient to prevent
the wrapping from peeling away from package. Failure to do so could
result in removal of the tape and ineffective sterilization.
1)
Wrap the tray containing the components.
2)
Apply the tape lengthwise across the three sides of the container.
1-3-3-2-4. Parameter
WARNING
Contact Johnson & Johnson regarding directions for use of STERRAD
system.
Set the tray containing the components into the sterilization chamber.
®
STERRAD
NX™:
Select "Advanced" cycle and operate the sterilizer according to the instructions supplied
®
with the STERRAD
NX™.
®
NX™ system.
– 49 –
®
NX™